Cross-calibration of multi-frequency bioelectrical impedance analysis with eight-point tactile electrodes and dual-energy X-ray absorptiometry for assessment of body composition in healthy children aged 6-18 years.

机译:八点触觉电极和双能X射线吸收光度法对多频生物电阻抗分析的交叉校准,以评估6-18岁健康儿童的身体成分。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: In diagnosis and treatment of obesity, body composition analysis including percent body fat (%BF) is useful in the clinical setting. Because bioelectrical impedance analysis (BIA) could be used quickly, easily and was non-invasive in clinical setting, the purpose of the present study was to evaluate the usefulness of multi-frequency BIA with eight-point tactile electrodes (MF-BIA8; InBody 720, Biospace) compared with dual-energy X-ray absorptiometry (DXA) in healthy children and adolescents. METHODS: A total of 166 children and adolescents under 18 (male, n = 86; female, n = 80) were recruited. Height, weight, body mass index (BMI) and Tanner stage were measured for each subject. The body composition such as fat-free mass (FFM), fat mass (FM), and %BF was measured on BIA and DXA and compared. RESULTS: On linear regression analysis, DXA FFM = 1.006(BIA FFM) + 0.554, R(2) = 0.99 and the standard error of the estimate (SEE) was 1.16 kg; DXA FM = 0.971(BIA FM) - 0.596, R(2) = 0.93; SEE, 1.34 kg; and DXA %BF = 0.940(BIA %BF) - 1.026, R(2) = 0.858; SEE, 3.03%. Limit of agreement in FFM, FM, and %BF was 0.7 +/- 2.3 kg, -0.9 +/- 2.9 kg and -2.2 +/- 6.1%, respectively. CONCLUSIONS: Although the %BF was not interchangeable with DXA, MF-BIA8 (InBody 720; Biospace) could be used to measure body composition of children and adolescents in the clinical field because of its high precision.
机译:背景:在肥胖的诊断和治疗中,包括身体脂肪百分比(%BF)在内的身体成分分析在临床环境中很有用。由于生物电阻抗分析(BIA)可以快速,轻松地在临床环境中使用,并且是非侵入性的,因此本研究的目的是评估带有八点触觉电极的多频BIA(MF-BIA8; InBody 720(Biospace)与健康儿童和青少年的双能X线吸收法(DXA)进行比较。方法:总共招募了166名18岁以下的儿童和青少年(男,n = 86;女,n = 80)。对每个受试者测量身高,体重,体重指数(BMI)和坦纳阶段。在BIA和DXA上测量身体成分,例如无脂肪量(FFM),脂肪量(FM)和%BF,并进行比较。结果:在线性回归分析中,DXA FFM = 1.006(BIA FFM)+ 0.554,R(2)= 0.99,估计的标准误差(SEE)为1.16 kg; DXA FM = 0.971(BIA FM)-0.596,R(2)= 0.93; SEE,1.34公斤;并且DXA%BF = 0.940(BIA%BF)-1.026,R(2)= 0.858;看到3.03%。 FFM,FM和%BF的一致性限制分别为0.7 +/- 2.3千克,-0.9 +/- 2.9千克和-2.2 +/- 6.1%。结论:尽管%BF与DXA不能互换,但MF-BIA8(InBody 720; Biospace)由于其高精度而可用于测量儿童和青少年的身体成分。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号