首页> 外文期刊>Pediatric Pulmonology >Increased incidence of sighs (augmented inspiratory efforts) during synchronized intermittent mandatory ventilation (SIMV) in preterm neonates.
【24h】

Increased incidence of sighs (augmented inspiratory efforts) during synchronized intermittent mandatory ventilation (SIMV) in preterm neonates.

机译:早产新生儿同步间歇性强制通气(SIMV)期间叹气(加大吸气力度)的发生率增加。

获取原文
获取原文并翻译 | 示例
       

摘要

A reflex resulting in a deep, sigh-like inspiratory effort (augmented breath) is frequently triggered during synchronized mechanical ventilation in preterm infants. We studied the incidence of augmented inspiratory efforts and their effect on ventilation and lung compliance during conventional IMV and synchronized IMV (SIMV) in 15 preterm neonates (GA 26.7 +/- 1.5 wks (mean +/- SD), BW 925 +/- 222 g, age 1-8 days). Augmentation of spontaneous inspiratory effort was defined as an esophageal pressure deflection occurring coincident with a synchronized mechanical breath and exceeding the previous unassisted spontaneous effort by more than 50%. The incidence of augmented breaths was higher during SIMV (11.1 +/- 7.7%; P < 0.01) than during conventional IMV (5.1 +/- 6.1%). However, when the synchronized breaths were triggered late (200-300 msec) after the onset of inspiration, augmented breaths occurred no more frequently than during conventional IMV (6.0 +/- 4.7%). The incidence of augmented breaths correlated inversely with dynamic lung compliance (P = 0.014), but was not significantly influenced by a change in PEEP. Although inspiratory effort increased nearly three times during the augmented breaths, tidal volume increased only 12%. The change in tidal volume was limited because the augmented effort reached its maximal negativity only approximately 500 ms after the beginning of the synchronized, mechanical breath and at a time when the mechanical breath had already ended. For this reason the augmented effort did not contribute significantly to minute ventilation, but only prolonged inspiration. Dynamic lung compliance did not change significantly after an augmented breath. The results indicate that augmented inspiratory efforts are more common in preterm neonates ventilated with SIMV than with conventional IMV, but do not contribute significantly to ventilation.
机译:在早产儿进行同步机械通气时,经常会触发反射,引起深沉的叹息状吸气动作(增强的呼吸)。我们研究了15名早产儿(GA 26.7 +/- 1.5 wks(平均+/- SD),BW 925 +/-)在常规IMV和同步IMV(SIMV)期间增加吸气努力的发生率及其对通气和肺顺应性的影响222克,年龄1-8天)。自发吸力的增加被定义为食管压力偏转与同步机械呼吸同时发生,并且比先前的无助自发力超出了50%以上。 SIMV期间增强呼吸的发生率较高(11.1 +/- 7.7%; P <0.01),而传统IMV期间则较高(5.1 +/- 6.1%)。但是,当在吸气开始后较晚(200-300毫秒)触发同步呼吸时,增强呼吸发生的频率不会比传统IMV期间更高(6.0 +/- 4.7%)。呼气的增加与动态肺顺应性呈负相关(P = 0.014),但不受PEEP变化的显着影响。尽管在增强呼吸过程中吸气量增加了近三倍,但潮气量仅增加了12%。潮气量的变化是有限的,因为在同步的机械呼吸开始后以及机械呼吸已经结束时,增强的力量仅在大约500毫秒时达到其最大负值。因此,加大力度不会显着促进分钟通气,而只会延长呼吸时间。呼吸增强后,动态肺顺应性没有明显改变。结果表明,与常规IMV相比,SIMV通气的早产儿更常见于加大吸气努力,但对通气无明显贡献。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号