With the exponential growth of cardiovascular implantable electronic devices (CIEDs) in pediatric patients, a new method of long-term surveillance, remote monitoring (RM), has become the standard of care. The purpose of this study was to determine the usefulness of RM as a monitoring tool in the pediatric population. A retrospective review was performed of 198 patients at the University of Iowa Children's Hospital who had CIEDs. Data transmitted by RM were analyzed. The following data were examined: patient demographics; median interval between transmissions; detection of adverse events requiring corrective measures, including detection of lead failure; detection of arrhythmias and device malfunctions independent of symptoms; time gained in the detection of events using RM versus standard practice; the validity of RM; and the impact of RM on data management. Of 198 patients, 162 submitted 615 RM transmissions. The median time between remote transmissions was 91 days. Of 615 total transmissions, 16 % had true adverse events with 11 % prompting clinical intervention. Of those events requiring clinical response, 61 % of patients reported symptoms. The median interval between last follow-up and occurrence of events detected by RM was 46 days, representing a gain of 134 days for patients followed-up at 6-month intervals and 44 days for patients followed-up at 3 month-intervals. The sensitivity and specificity of RM were found to be 99 and 72 %, respectively. The positive and negative predictive values were found to be 41 and 99 %, respectively. RM allows for early identification of arrhythmias and device malfunctions, thus prompting earlier corrective measures and improving care and safety in pediatric patients.
机译:随着儿科患者中可植入心血管电子设备(CIED)的呈指数增长,一种新的长期监视,远程监视(RM)方法已成为护理的标准。这项研究的目的是确定RM作为儿科人群监测工具的有用性。对爱荷华大学儿童医院的198名有CIED的患者进行了回顾性审查。对RM传输的数据进行了分析。检查了以下数据:患者人口统计数据;传输之间的中间间隔;检测需要采取纠正措施的不良事件,包括检测铅失效;独立于症状的心律不齐和设备故障的检测;使用RM与标准做法相比,在事件检测中获得的时间; RM的有效性;以及RM对数据管理的影响。在198位患者中,有162位患者提交了615次RM传播。远程传输之间的平均时间为91天。在615例总传播中,有16%发生真正的不良事件,其中11%促使临床干预。在那些需要临床反应的事件中,有61%的患者报告了症状。上次随访与RM检测到的事件发生之间的中位间隔为46天,对于以6个月为间隔进行随访的患者,增加了134天;对于以3个月为间隔进行随访的患者,则增加了44天。发现RM的敏感性和特异性分别为99%和72%。发现阳性和阴性预测值分别为41%和99%。 RM可以及早发现心律不齐和设备故障,从而提早采取纠正措施并改善儿科患者的护理和安全性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号