首页> 外文期刊>Spectrochimica Acta, Part B. Atomic Spectroscopy >Determination of Cu and Mn in seawater by graphite furnace atomic absorption spectrometry with the use of hydrofluoric acid as a chemical modifier
【24h】

Determination of Cu and Mn in seawater by graphite furnace atomic absorption spectrometry with the use of hydrofluoric acid as a chemical modifier

机译:氢氟酸为化学改性剂的石墨炉原子吸收光谱法测定海水中的铜和锰

获取原文
获取原文并翻译 | 示例

摘要

The addition of hydrogen fluoride to a seawater sample in a graphite tube atomizer makes possible the removal of the major matrix components in an optimized pre-heating step, thus a near interference free determination of copper and manganese trace metals is achieved. The optimization studies were supported by a priori calculation of limit of detection (LOD) data. For optimized conditions, the experimentally determined LOD values (3sigma, n = 20) were 0.11 and 0.03 mug l(-1) for Cu and Mn, respectively, for 20 mul applied seawater volume. With the use of calibration with standard addition, determined concentrations of Cu and Mn in CASS-2 seawater and in SLEW-1 estuarine water were in good agreement with certified values at the 95% confidence level (n = 10). (C) 2002 Elsevier Science B.V. All rights reserved. [References: 29]
机译:在石墨管雾化器中将氟化氢添加到海水样品中,可以在优化的预热步骤中去除主要基质成分,从而实现了铜和锰微量金属的几乎无干扰的测定。优化研究得到了检测限(LOD)数据的先验计算的支持。对于最佳条件,实验确定的LOD值(3sigma,n = 20)对于20 mul施加的海水量,对于Cu和Mn分别为0.11和0.03马克杯l(-1)。通过使用标准添加物进行校准,确定的CASS-2海水和SLEW-1河口水中Cu和Mn的浓度与95%置信水平(n = 10)的认证值高度吻合。 (C)2002 Elsevier Science B.V.保留所有权利。 [参考:29]