Determination of Boron in Blood, Urine and Bone by Electrothermal Atomic Absorption Spectrometry Using Zirconium and Citric Acid as Modifiers

机译:锆和柠檬酸为改性剂的电热原子吸收光谱法测定血液,尿液和骨中的硼

获取原文
获取原文并翻译 | 示例

摘要

A comparative study of various potential chemical modifiers (Au, Ba, Be, Ca,Cr, Ir, La, Lu, Mg, Ni, Pd,Pt,Rh, Ru, Sr, V, W, and Zr) and different 'coating' treatment (Zr, W, and W + Rh) of the pyrolytic graphite platform of a longitudinally heated graphite tube atomizer for thermalstabilization and determination of boron was undertaken. The use of Au, Ba, Be, Cr, Ir, Pt, Rh, Sr and V as modifiers, and W + Rh coating produced erratic, and noisy signals, while the addition of La, Ni and Pd as modifiers, and the W coating had positive effects, biu with too high background absorption signals, rendering their use unsuitable for boron determination even in aqueous solutions. The atomic absorption signal for boron was increased and stabilized when the platform was coated with Zr, and by the addition of Ca, Mg, Lu, W or Zr as modifiers. Only the addition of 10 #mu#g of Zr as a modifier onto Zr-treated platforms allowed the use of a higher pyrolysis temperature without analyte losses. The memory effect was minimized by incorporating a cleaning step with 10 #mu#l of 50 g l~(-1) NH_4F HF after every three boron measurements. The addition of 10 #mu#l of 15 g l~(-1) citric acid together with Zr onto Zr-treated platforms significantly improved the characteristic mass to m_0 = 282 pg, which is adequate for biological samples such as urine and bone, although the sensitivity was still inadequate for the detemrination of boron in blood of subjects without supplementary diet. Under optimized conditions, the detection limit (3#sigma#) was 60 #mu#g l~(-1). The amount of boron found in whole blood, urine and femur head samples from patients with osteoporosis was in agreement with values previously reported in the literature.
机译:各种潜在化学改性剂(Au,Ba,Be,Ca,Cr,Ir,La,Lu,Mg,Ni,Pd,Pt,Rh,Ru,Sr,V,W和Zr)和不同涂层的比较研究进行了纵向加热的石墨管雾化器的热解石墨平台的热处理(Zr,W和W + Rh),以进行热稳定和硼的测定。使用Au,Ba,Be,Cr,Ir,Pt,Rh,Sr和V作为修饰剂,而W + Rh涂层产生不稳定且嘈杂的信号,而添加La,Ni和Pd作为修饰剂,而W涂层具有积极作用,背景吸收信号过高,甚至在水溶液中也不适用于硼的测定。当平台涂有Zr并添加Ca,Mg,Lu,W或Zr作为改性剂时,硼的原子吸收信号增加并稳定。只有在Zr处理过的平台上添加10#mu#g Zr作为改性剂,才可以使用更高的热解温度而不会造成分析物损失。每三个硼测量后,通过加入10 g / l的50 g l〜(-1)NH_4F HF清洗步骤,将记忆效应降至最低。在Zr处理的平台上添加10#mu#l的15 gl〜(-1)柠檬酸和Zr可以将特性质量显着提高到m_0 = 282 pg,尽管对于尿液和骨骼等生物样品来说还是足够的对于没有补充饮食的受试者血液中的硼测定,敏感性仍然不足。在最佳条件下,检出限(3#sigma#)为60#mu#g l〜(-1)。骨质疏松症患者的全血,尿液和股骨头样品中发现的硼含量与文献中先前报道的值一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号