The risk of foraminal violation and nerve root impingement after anterior placement of lumbar interbody fusion cages.

机译:腰椎椎间融合器放置前,会发生椎间孔侵犯和神经根受侵的风险。

获取原文
获取原文并翻译 | 示例
           

摘要

STUDY DESIGN: Three groups of six embalmed cadaver spines underwent placement of lumbar interbody fusion cages centered either at midline, 10% lateral of midline, or 20% lateral of midline. The spines were evaluated for evidence of neuroforamen violation or nerve root impingement. OBJECTIVES: To determine the potential for foraminal violation or nerve root impingement after correct placement and lateral misplacement of lumbar interbody fusion cages. SUMMARY OF BACKGROUND DATA: Radicular symptoms after anterior cage placement have raised some concern about the potential for inadvertent device-related foraminal violation not adequately appreciated by intraoperative fluoroscopy. METHODS: Preoperative computed tomography scanning and plain radiography was used to measure endplate dimensions at L4-L5 and to template the appropriately sized interbody fusion cages. The cadaveric specimens were randomly divided into three groups of six (Groups I-III) and instrumented at L4-L5 either at midline (I) or 10% (II) or 20% (III) lateral of midline. Postoperative computed tomography and plain radiography was evaluated for evidence of neuroforamen violation, followed by dissection of the specimens. RESULTS: Foraminal violation occurred in one of six spines in group II (10% off midline) and in three of six spines in group III (20% off midline). Two of the three cadavers in group III with foraminal violation also were noted to have nerve root abutment on computed tomography scans and spinal dissection. CONCLUSIONS: Excessive lateral placement of lumbar interbody fusion cages may result in foraminal violation and possible nerve encroachment. The "safe zone" for centering the cages extends approximately 5 mm on either side of midline.
机译:研究设计:将三组六个防腐的尸体脊椎置于腰椎椎间融合器,定位于中线,中线外侧10%或中线外侧20%。评估脊柱是否有侵犯神经孔或侵犯神经根的证据。目的:确定正确放置腰椎椎间融合器并发生侧向错位后,是否有孔眼侵犯或神经根受冲击的可能性。背景数据摘要:前笼放置后的根性症状引起了人们对术中透视无法充分意识到与器械相关的椎间孔侵犯的潜在可能性的担忧。方法:术前使用计算机断层扫描和X线平片照相术来测量L4-L5的终板尺寸,并对适当尺寸的椎间融合器进行模板化。将尸体标本随机分为三组,每组六个(I-III组),并在中线外侧(I)或10%(II)或20%(III)的L4-L5处进行测量。评估术后计算机断层扫描和X线平片检查是否有神经孔穿孔的证据,然后解剖标本。结果:第二组的六个棘突之一(中线偏离10%)和第三组的六个棘突中的三个棘轮(偏离中线20%)发生了椎间孔侵犯。在第三组尸体中,有孔眼侵犯的三具尸体中有两具在计算机断层扫描和脊柱解剖学上具有神经根基台。结论:腰椎椎间融合器横向放置过多可能会导致椎间孔侵犯和可能的神经侵犯。使笼子居中的“安全区”在中线两侧延伸约5毫米。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号