The migration of femoral components after total hip replacement surgery: accuracy and precision of software-aided measurements.

机译:全髋关节置换手术后股骨组件的迁移:软件辅助测量的准确性和精确性。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE. To assess the accuracy and precision of a software-aided system to measure migration of femoral components after total hip replacement (THR) on digitised radiographs. DESIGN AND PATIENTS. Subsidence and varus-valgus tilt of THR stems within the femur were measured on digitised anteroposterior pelvic radiographs. The measuring software (UMA, GEMED, Germany) relies on bony landmarks and comparability parameters of two consecutive radiographs. Its accuracy and precision were calculated by comparing it with the gold standard in migration measurements, radiostereometric analysis (RSA). Radiographs and corresponding RSA measurements were performed in 60 patients (38-69 years) following cementless THR surgery. RESULTS AND CONCLUSIONS. The UMA software measured the subsidence of the stems with an accuracy of +/-2.5 mm and varus-valgus tilt with an accuracy of +/-1.8 degrees (95% confidence interval). A good interobserver and intraobserver reliability was calculated with Cronbach's alpha ranging from0.86 to 0.97. Measuring the subsidence of THR stems within the femur is an important parameter in the diagnosis of implant loosening. Software systems such as UMA improve the accuracy of migration measurements and are easy to use on routinely performed radiographs of operated hip joints.
机译:目的。评估数字化X线片上全髋关节置换(THR)后,用于测量股骨组件迁移的软件辅助系统的准确性和精度。设计和患者。用数字化前后骨盆X光片测量股骨内THR茎的下沉和内翻-外翻倾斜度。测量软件(UMA,GEMED,德国)依赖于骨标和两个连续X射线照片的可比性参数。通过与迁移测量,放射立体分析(RSA)中的金标准进行比较,计算出其准确性和精确度。非骨水泥性THR手术后对60例患者(38-69岁)进行了X线照片和相应的RSA测量。结果与结论。 UMA软件以+/- 2.5 mm的精度测量茎杆的沉降,以+/- 1.8度(95%置信区间)的精度测量内翻-外翻倾斜度。计算得出良好的观察者间和观察者内可靠性,Cronbach的alpha值在0.86至0.97之间。测量股骨内THR茎下沉是诊断植入物松动的重要参数。诸如UMA之类的软件系统提高了迁移测量的准确性,并且易于在操作的髋关节的常规X光片上使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号