Implantable accelerometer system for the determination of blood pressure using reflected wave transit time☆

机译:植入式加速度计系统,利用反射波的传播时间来确定血压☆

获取原文
获取原文并翻译 | 示例
           

摘要

A sensor system for continuous monitoring of blood pressure using an acceleration sensor implanted on an artery using minimally invasive techniques is described. The sensor relies on measurement of the reflected wave transit time (RWTT). This implantable system is fabricated on a flexible substrate using 2 mm × 2 mm acceleration sensors and a telemetric unit for transmission of the data. In vivo experiments show that the RWTT can be reliably determined from arterial acceleration signals. RWTT and systolic blood pressure are shown to be strongly coupled, with a correlation coefficient of 0.96, as determined from measurement of 1800 pulses with a mean deviation of the blood pressure of only 4.3%. The system was implanted in an animal and was able to telemetrically transmit acceleration plethysmographs with high quality out of the awake animal.
机译:描述了一种使用微创技术通过植入在动脉上的加速度传感器连续监测血压的传感器系统。传感器依赖于反射波传播时间(RWTT)的测量。该可植入系统使用2 mm×2 mm加速度传感器和用于传输数据的遥测单元制造在柔性基板上。体内实验表明,可以从动脉加速度信号可靠地确定RWTT。 RWTT和收缩压显示出强耦合,相关系数为0.96,这是通过测量1800个脉搏而测得的,血压的平均偏差仅为4.3%。该系统被植入动物体内,并能够从清醒的动物中遥测发送高质量的加速度体积描记器。