首页> 外文期刊>Otology and neurotology: official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology >Intraoperative auditory steady state response measurements during Vibrant Soundbridge middle ear implantation in patients with mixed hearing loss: preliminary results.
【24h】

Intraoperative auditory steady state response measurements during Vibrant Soundbridge middle ear implantation in patients with mixed hearing loss: preliminary results.

机译:混合性听力损失患者在充满声桥中耳植入过程中的术中听觉稳态反应测量:初步结果。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVE: To optimize intraoperatively the coupling of the floating mass transducer (FMT) of the Vibrant Soundbridge middle ear implant to the round or oval cochlear window in patients with mixed hearing loss. STUDY DESIGN: Intraoperative measurement of objective hearing thresholds using auditory steady state responses (ASSRs). SETTING: Radboud University Nijmegen Medical Centre, tertiary referral hospital. PATIENTS: Four individuals with mixed hearing loss and, at least no incus, in need for a middle ear implant. INTERVENTION: Surgical placement of the Vibrant Soundbridge. ASSR thresholds were measured intraoperatively, whereas FMT coupling to the cochlea was manipulated to find the most effective coupling of the FMT. MAIN OUTCOME MEASURE: Differences in ASSR thresholds between different FMT coupling options within patients. RESULTS: With ASSR, we assessed placement of the FMT in the round window niche, loosely or tightly packed in the niche; creation of a mobile window in case of a fixed stapes footplate; and FMT position coupled to the stapes that afforded vibration in the natural vibration direction or perpendicular to it. Furthermore, test-retest variations in ASSR thresholds were studied. It was shown that differences in ASSR thresholds could be detected, while manipulating the FMT couplings, which were statistically significant. CONCLUSION: Intraoperative ASSR measurement is a good method to study different positions of the FMT and to determine the best position of the FMT for a patient.
机译:目的:为混合听力损失患者在术中优化振动声桥中耳植入物的漂浮质量传感器(FMT)与圆形或椭圆形耳蜗窗的耦合。研究设计:术中使用听觉稳态反应(ASSR)进行客观听力阈值的测量。地点:拉德布德大学奈梅亨医学中心,三级转诊医院。患者:四个有混合性听力损失且至少没有骨cus的患者需要中耳植入物。干预:充满活力的声桥的手术放置。术中测量ASSR阈值,而操纵FMT与耳蜗的耦合以找到最有效的FMT耦合。主要观察指标:患者体内不同FMT偶联方案之间的ASSR阈值差异。结果:通过ASSR,我们评估了FMT在松散或紧密包装在小生境中的圆形窗口小生境中的放置;在固定骨踏板的情况下创建活动窗口; FMT位置与在自然振动方向或垂直于自然振动方向振动的骨相连。此外,研究了ASSR阈值的重测差异。结果表明,在操纵FMT耦合时,可以检测到ASSR阈值的差异,这在统计上是显着的。结论:术中ASSR测量是研究FMT不同位置并确定患者最佳FMT位置的好方法。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号