首页> 外文期刊>Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology and Endodontics >Niobium filtration of conventional and high-frequency x-ray generator beams for intraoral radiography. Effects on absorbed doses, image density and contrast, and photon spectra.
【24h】

Niobium filtration of conventional and high-frequency x-ray generator beams for intraoral radiography. Effects on absorbed doses, image density and contrast, and photon spectra.

机译:用于口腔内放射照相的常规和高频X射线发生器束的铌过滤。对吸收剂量,图像密度和对比度以及光子光谱的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

We have studied the effects of niobium beam filtration on absorbed doses, on image density and contrast, and on photon spectra with conventional and high-frequency dental x-ray generators. Added niobium reduced entry and superficial absorbed doses in periapical radiography by 9% to 40% with film and digital image receptors, decreased the radiation necessary to produce a given image density on E-speed film and reduced image contrast on D- and E-speed films. As shown by increased half-value layers for aluminum, titanium, and copper and by pulse-height analyses of beam spectra, niobium increased average beam energy by 6% to 19%. Despite the benefits of adding niobium on patient dose reduction and on narrowing the beams' energy spectra, the beam can be overhardened. Adding niobium, therefore, strikes the best balance between radiation dose reduction and beam attenuation, with its risks of increased exposure times, motion blur, and diminished image contrast, when it is used at modest thicknesses (30 microns) and at lower kVp (70) settings.
机译:我们使用常规和高频牙科X射线发生器研究了铌束过滤对吸收剂量,图像密度和对比度以及光子光谱的影响。添加了铌后,利用胶片和数字图像接收器,可以使根尖X线摄影中的进入和表面吸收剂量减少9%至40%,减少了在E速胶片上产生给定图像密度所需的辐射,并降低了D速和E速成像的对比度电影。如铝,钛和铜的半值层增加以及光束光谱的脉冲高度分析所示,铌将平均光束能量提高了6%至19%。尽管添加铌有益于减少患者剂量并缩小束的能谱,但束可能会过硬。因此,添加铌在中等厚度(30微米)和较低kVp(70)下使用时,会增加曝光时间,运动模糊和图像对比度降低,从而在减少辐射剂量和减少光束衰减之间达到最佳平衡。 )设置。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号