首页> 外文期刊>Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology and Endodontics >Validation of the Bitestrip versus polysomnography in the diagnosis of patients with a clinical history of sleep bruxism
【24h】

Validation of the Bitestrip versus polysomnography in the diagnosis of patients with a clinical history of sleep bruxism

机译:Bitestrip与多导睡眠图在睡眠磨牙症临床病史诊断中的验证

获取原文
获取原文并翻译 | 示例
           

摘要

Objective. This validation study compared an electromyography (EMC) device, Bitestrip, versus polysomnography in the diagnosis of sleep bruxism (SB). Study design. Forty-nine consecutive patients (32 women and 17 men, 41.2 ± 12.9 years old) with a clinical history of SB were included in the study from the Orofacial Pain Clinic, Faculty of Dentistry, Pontifical Catholic University of Rio Grande do Sul in Brazil. Results. Initially, we compared an SB positiveegative test result for both systems (agreement 87.8%, sensitivity 84.2%, positive predictive value 100%, and crude kappa 0.71). We then compared a 4-scale test (no, light, moderate, or severe bruxism) between the 2 methods (agreement 80.27%, weighted kappa 0.51, and Kendall W coefficient 0.575). Conclusions. The Bitestrip can be considered as a moderate screening method for the diagnosis of SB, because it is more precise in detecting the presence or absence of SB but less precise in detecting its intensity.
机译:目的。该验证研究比较了肌电图(EMC)设备Bitestrip与多导睡眠图在睡眠磨牙症(SB)的诊断中的作用。学习规划。来自巴西里奥格兰德多宗天主教大学牙学院口腔颌面疼痛诊所的这项研究包括了49例具有SB临床病史的连续患者(32例女性和17例男性,41.2±12.9岁)。结果。最初,我们比较了两个系统的SB阳性/阴性测试结果(协议87.8%,敏感性84.2%,阳性预测值100%和粗kappa 0.71)。然后,我们比较了两种方法(协议80.27%,加权kappa 0.51和Kendall W系数0.575)之间的4等级测试(无,轻度,中度或重度磨牙症)。结论。 Bitestrip可被认为是诊断SB的中度筛选方法,因为它在检测SB存在与否方面更精确,而在检测其强度方面却不那么精确。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号