Reliability of laser Doppler flowmetry in a 2-probe assessment of pulpal blood flow.

机译:激光多普勒血流仪在牙髓血流2探头评估中的可靠性。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: The reliability of using 2 probes with laser Doppler signals when adjacent teeth are being measured simultaneously to determine pulpal blood flow is unknown. The purpose of this study was to determine whether 2 probes are more reliable than 1 in a single-tool assessment. STUDY DESIGN: Tooth pulp vitality was studied in 19 adults through use of laser Doppler flowmetry tests. In each subject, testing was carried out on 2 successive occasions with 2 probes positioned on the maxillary central incisors. RESULTS: Significant mean differences of 31% for blood flux and 96% for concentration were found between the 2 probes, although they shared equal coefficients of variation. The reproducibility for each probe was found to be consistent, and the probes were highly correlated with each other. Flux and concentration, however, were not systematically correlated. CONCLUSIONS: Simultaneous measurements with 2 probes were clearly more reliable. The necessity for a calibration control was evident.
机译:目的:当同时测量相邻牙齿以确定牙髓血流时,使用两个带有激光多普勒信号的探头的可靠性尚不清楚。这项研究的目的是确定在单工具评估中2支探针是否比1支更可靠。研究设计:通过使用激光多普勒血流仪测试对19名成年人的牙髓活力进行了研究。在每个受试者中,连续2次在上颌中切牙上放置2个探头进行测试。结果:尽管两个探针具有相同的变异系数,但在两个探针之间发现显着的平均血流差异为31%,浓度为96%。发现每种探针的重现性是一致的,并且探针彼此高度相关。通量和浓度,但是,没有系统地相关。结论:用2个探头同时测量显然更可靠。校准控制的必要性显而易见。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号