首页> 外文期刊>Ophthalmology >Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects.
【24h】

Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects.

机译:动态轮廓眼压测量法和戈德曼压平眼压测量法在非裔美国人受试者中的比较。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To evaluate the relationship between intraocular pressure (IOP) measurements obtained by dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in African Americans and to assess whether these measures were influenced by ocular parameters including corneal thickness, corneal curvature, and axial length. DESIGN: Observational clinical study. PARTICIPANTS: The study included 176 eyes of 94 African Americans. METHODS: All participants underwent IOP evaluation with DCT and GAT, as well as measurements of central corneal thickness (CCT), corneal curvature, corneal astigmatism, spherical equivalent, and axial length. MAIN OUTCOME MEASURES: Univariate and multiple regression analyses were used to evaluate the associations between IOP (as measured with DCT and GAT) and CCT, corneal curvature, corneal astigmatism, spherical equivalent, axial length, and age. Bland-Altman plots were used to evaluate the agreement between IOP measurements obtained by DCT and GAT. RESULTS: In multiple regression analysis, GAT IOP measurements were significantly associated with CCT (P<0.001) and age (P = 0.003), whereas DCT IOP measurements were associated only with age (P = 0.027). The difference between DCT and GAT IOP measurements was significantly influenced by corneal thickness (P<0.001) and ocular pulse amplitude (OPA; P = 0.004). Patients with thick corneas tended to have higher GAT IOP measurements compared with DCT, whereas in patients with thin corneas, GAT IOP measurements tended to be lower than DCT measurements. Larger values of OPA were associated with a decrease in the difference between DCT and GAT measurements, whereas smaller values of OPA were associated with an increase in the difference between DCT and GAT measurements. CONCLUSIONS: Dynamic contour tonometry measurements in African Americans seem to provide an estimate of IOP that is less influenced by corneal properties than those provided by GAT.
机译:目的:评估非裔美国人通过动态轮廓眼压计(DCT)和戈德曼压平眼压计(GAT)获得的眼内压(IOP)测量值之间的关系,并评估这些测量值是否受到包括角膜厚度,角膜曲率和轴向长度。设计:观察性临床研究。参与者:该研究包括94名非裔美国人的176眼。方法:所有参与者均接受DCT和GAT的IOP评估,并测量中央角膜厚度(CCT),角膜曲率,角膜散光,等效球体和轴向长度。主要观察指标:单因素和多元回归分析用于评估眼压(用DCT和GAT测量)与CCT,角膜曲率,角膜散光,当量球体,轴向长度和年龄之间的关联。使用Bland-Altman图来评估DCT和GAT获得的IOP测量值之间的一致性。结果:在多元回归分析中,GAT IOP测量值与CCT(P <0.001)和年龄(P = 0.003)显着相关,而DCT IOP测量值仅与年龄(P = 0.027)相关。 DCT和GAT IOP测量值之间的差异受到角膜厚度(P <0.001)和眼动幅度(OPA; P = 0.004)的显着影响。与DCT相比,角膜较厚的患者往往具有更高的GAT IOP测量值,而角膜较薄的患者中,GAT IOP测量结果往往低于DCT测量值。较大的OPA值与DCT和GAT测量值之间的差异减小相关,而较小的OPA与DCT和GAT测量值之间的差异增加相关。结论:非洲裔美国人的动态轮廓眼压测量法似乎提供了IOP的估计值,该估计值受角膜特性的影响小于GAT提供的值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号