Comparison of spherical equivalent refraction and astigmatism measured with three different models of autorefractors.

机译:使用三种不同型号的自动折射仪测量的球面等效折射和像散的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The purpose of this study was to compare refractions measured with three different autorefractors. METHODS: The refractive error of each eye of 50 adults aged 17 to 59 years (mean, 30.5 years) was measured without cycloplegia using the Canon R-1 and two newer instruments, the Grand Seiko WR-5100K and the Nidek ARK 700-A. For the first two, an isolated line of 20/100 letters on an ETDRS chart at 4.0 m served as a target, whereas for the Nidek, the subject looked at a picture of a balloon in the instrument. Five readings were taken for each eye, and the data (sphere, negative cylinder power, and axis) were analyzed using Fourier decomposition of the power profile. Each reading was broken down into the spherical equivalent (M) and two Jackson crossed-cylinder vectors, J0 and J45. Right-eye results are reported. RESULTS: The mean spherical equivalent refraction measured by the Canon R-1 was -2.44 D. Measurements from the Grand Seiko were more hyperopic (mean M, -2.01 D), whereas those from the Nidek were more myopic (mean M, -2.66 D). Correlation of M for each pair of autorefractors was 0.99. For J0, the Canon was more minus than the other two instruments by 0.15 D compared with the Nidek and 0.13 D with the Grand Seiko, and on this component, the correlation of the Canon with each of the other two was 0.87. Mean J0 values for the Nidek and Grand Seiko were similar, 0.05 D and 0.03 D, respectively, and mean J45 values were 0.04 D for both instruments. Correlations were 0.97 between these two autorefractors for each of the two components. J45 measured by the Canon was more positive than the other two by 0.06 D. For J45, the correlation of the Canon with the Grand Seiko was 0.40 and with the Nidek was 0.38. In 92% of the eyes, the absolute difference in cylinder power between the Grand Seiko and the Nidek was < or =0.25 D. Only 42% of the differences between the Canon and the Grand Seiko and 40% of the differences between the Canon and the Nidek were this small. CONCLUSIONS: The Canon provided more myopic readings than the Grand Seiko and more hyperopic readings than the Nidek. The Canon measured more astigmatism that did not correlate well with the other instruments, whereas measurements of astigmatism taken by the Nidek and the Grand Seiko showed good agreement. If the Grand Seiko is used in place of the older Canon, the differences in spherical equivalent and astigmatism must be considered.
机译:目的:本研究的目的是比较使用三种不同的自动折射仪测量的折射。方法:使用佳能R-1和两款新仪器Grand Seiko WR-5100K和Nidek ARK 700-A测量了50例年龄在17至59岁(平均30.5岁)的成年人的眼睛的屈光不正,没有睫状肌麻痹。 。对于前两个,在4.0 m处的ETDRS图表上的20/100个字母的隔离线作为目标,而对于Nidek,对象则看着仪器中气球的图像。每只眼睛获取五个读数,并使用功效曲线的傅立叶分解分析数据(球面,负圆柱度和轴)。每个读数都分解为等效的球面(M)和两个杰克逊交叉圆柱向量J0和J45。报告右眼结果。结果:佳能R-1测得的平均球面等效屈光度为-2.44D。GrandSeiko的测光度数为远视(平均M,-2.01 D),而Nidek的测光度数为近视(平均M,-2.66)。 D)。每对自动验光仪的M相关系数为0.99。对于J0,佳能比Nidek少0.15 D,佳能比Grand Seiko小0.13 D,是佳能与其他两款乐器的差值,并且在该组件上,佳能与其他两款乐器的相关性是0.87。 Nidek和Grand Seiko的平均J0值相似,分别为0.05 D和0.03 D,两种仪器的平均J45值为0.04D。对于这两个分量,这两个自动折射仪之间的相关性均为0.97。佳能测得的J45比其他两个测得的值高0.06D。对于J45,佳能与Grand Seiko的相关性为0.40,与Nidek的相关性为0.38。在92%的眼睛中,Grand Seiko和Nidek之间的柱面度绝对差为<或= 0.25D。佳能和Grand Seiko之间只有42%的差异,Canon和Grand Seiko之间只有40%的差异Nidek很小。结论:佳能提供了比Grand Seiko更多的近视读数,以及比Nidek更多的远视读数。佳能测得的散光与其他仪器的相关性不佳,而尼德克和精工制造的散光的测量显示出很好的一致性。如果使用Grand Seiko代替较旧的Canon,则必须考虑等效球差和像散的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号