We evaluated the quality of imaging posterior chamber intraocular lenses (PC IOL) by Scheimpflug photography. We took photographs of 190 rigid or foldable PC IOLs using the Scheimpflug system EAS-1000 from Nidek and provided digital image analysis. We found a good imaging of PC IOLs and the cristalline lens capsule, data obtained by digital image analysis were reliable and reproducible and Scheimpflug photography could be demonstrated as a sensitive tool to show any irregularities of IOL material. Thus, we found in foldable IOLs 'glistenings', which did not have any influence on functional results and we could distinguish them from lens damage. We propose to use Scheimpflug tests regularly for the examination of implanted new types of IOLs.
机译:我们通过Scheimpflug摄影评估了后房人工晶状体(PC IOL)成像的质量。我们使用Nidek的Scheimpflug系统EAS-1000拍摄了190个刚性或可折叠PC IOL的照片,并提供了数字图像分析。我们发现PC IOL和晶状体晶状体囊的成像良好,通过数字图像分析获得的数据可靠且可重现,Scheimpflug摄影可以作为一种敏感工具来显示IOL材料的任何不规则性。因此,我们在可折叠的IOL中发现了“闪闪发光”,这对功能结果没有任何影响,我们可以将它们与晶状体损伤区分开来。我们建议定期使用Scheimpflug测试来检查植入的新型IOL。