首页> 外文期刊>Research & reviews in electrochemistry >Determination of ultra trace of thiosemicarbazide by adsorptive stripping voltammetric method
【24h】

Determination of ultra trace of thiosemicarbazide by adsorptive stripping voltammetric method

机译:吸附溶出伏安法测定痕量硫代氨基脲。

获取原文
获取原文并翻译 | 示例
       

摘要

In the present work, an adsorptive stripping voltammetric method using a hanging mercury drop electrode (HMDE) was described in order to determine the ultra trace of thiosemicarbazide in different real samples. The method is based on accumulation of thiosemicarbazide on mercury electrode. The potential was scanned to the negative direction and the differential pulse stripping voltammograms were recorded. The instrumental and chemical parameters were optimized. The optimized conditions were obtained in pH of 10.0, accumulation potential of 0.00 mV, accumulation time of 60 s, scan rate of 80 mV/s and pulse height of 50 mV. The relationship between the peak current versus concentration was linear over the range of 0.50-100.00 ng/ml. The limits of detection were 0.03 ng/ml and the relative standard deviation at 5.00 and 50.00 ng/ml of thiosemicarbazide ion were obtained 2.1 and 1.7%, respectively (n = 8).
机译:在当前的工作中,描述了一种使用悬挂式汞滴电极(HMDE)的吸附溶出伏安法,以便测定不同真实样品中硫代氨基脲的超痕量。该方法基于硫代氨基脲在汞电极上的积累。将电位扫描到负方向,并记录差分脉冲溶出伏安图。仪器和化学参数进行了优化。在10.0的pH值,0.00 mV的累积电势,60 s的累积时间,80 mV / s的扫描速率和50 mV的脉冲高度下获得了最佳条件。峰值电流与浓度之间的关系在0.50-100.00 ng / ml的范围内呈线性关系。检出限为0.03 ng / ml,硫代氨基脲离子在5.00和50.00 ng / ml处的相对标准偏差分别为2.1和1.7%(n = 8)。