首页> 外文期刊>Resuscitation. >A novel method to decrease mattress compression during CPR using a mattress compression cover and a vacuum pump
【24h】

A novel method to decrease mattress compression during CPR using a mattress compression cover and a vacuum pump

机译:使用床垫压缩套和真空泵降低心肺复苏过程中床垫压缩的新颖方法

获取原文
获取原文并翻译 | 示例
       

摘要

Background: Mattress compression causes feedback devices to over-estimate the chest compression depth measurement during CPR. We propose a novel method to decrease the mattress compression using a vinyl cover. This mattress compression cover encloses the foam mattress and is compressed by a vacuum pump immediately prior to performing CPR. Methods: Nine CPR providers performed chest compressions on manikins placed on a conventional foam mattress on a bed frame (surface CONV), a backboard and foam mattress on a bed frame (surface BB), and a foam mattress, compressed with a vacuum pump, on a bed frame (surface VAC). Dual accelerometers were used to simultaneously measure the mattress compression and chest compression depths. Results: The mattress compression depth levels decreased from 14.9. mm (SD 1.4. mm) on surface CONV to 7.0. mm (SD 0.6. mm) on surface VAC (p<. 0.001) whereas 14.0. mm (SD 1.3. mm) on surface BB. The total compression depth was 65.4. mm (SD 3.8. mm) on surface CONV, and 58.3. mm (SD 3.0. mm) on surface VAC (p<. 0.001). Conclusion: Using a mattress compression cover and a vacuum pump appears to increase the rigidity of the mattress and allow for efficient chest compressions. This novel method could decrease the mattress compression depth and increase the efficiency of chest compression during CPR in hospitals.
机译:背景:床垫压缩会导致反馈设备在CPR期间过高估计胸部按压深度的测量值。我们提出了一种新颖的方法来减少使用乙烯基覆盖物的床垫压缩。该床垫压缩套包围泡沫床垫,并在执行CPR之前立即由真空泵压缩。方法:九名CPR提供者对放置在床架(表面CONV)上常规泡沫床垫上的人体模型,床架(BB表面)上的篮板和泡沫床垫,以及用真空泵压缩的泡沫床垫进行了胸部按压,在床架上(表面VAC)。双加速度计用于同时测量床垫的受压和胸部受压的深度。结果:床垫的压缩深度水平从14.9降低。表面CONV上的mm(SD 1.4。mm)至7.0。 VAC(p <.0.001)时为mm(SD 0.6。mm),而14.0。 BB面上的毫米(SD 1.3。毫米)。总压缩深度为65.4。表面CONV上的毫米(SD 3.8。mm)和58.3。 VAC(p <.0.001)上的mm(SD 3.0。mm)。结论:使用床垫压缩套和真空泵似乎可以增加床垫的刚度并可以有效地胸部按压。这种新颖的方法可以减少床垫的按压深度,并在医院进行心肺复苏时提高胸部按压的效率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号