首页> 外文期刊>Radiation Physics and Chemistry >Peak shifted properties of the 'low background NaI(Tl) detectors': An experimental study of response function behavior in different temperature and acquisition time
【24h】

Peak shifted properties of the 'low background NaI(Tl) detectors': An experimental study of response function behavior in different temperature and acquisition time

机译:“低背景NaI(Tl)检测器”的峰移特性:不同温度和采集时间下响应函数行为的实验研究

获取原文
获取原文并翻译 | 示例

摘要

Due to the necessity of using low background NaI detector in sensitive and accurate measurements, study on the response function variations in different conditions is very important. These types of detectors have different responses in various measurement conditions, including time, temperature and high voltage. In this study, the response function of 76 B 76 LB NaI (SCIONIX) in different conditions is discussed. According to the channel shifting in these detectors and its direct effect on degrading the resolution, the most convenient measurement condition for these detectors, is proposed. Finally, it is recommended that before long-time measurements a "waiting time" is needed to avoid the channel shifting effects. (C) 2016 Elsevier Ltd. All rights reserved.
机译:由于在灵敏而准确的测量中必须使用低背景NaI检测器,因此研究不同条件下的响应函数变化非常重要。这些类型的检测器在各种测量条件(包括时间,温度和高压)下具有不同的响应。在这项研究中,讨论了76 B 76 LB NaI(SCIONIX)在不同条件下的响应功能。根据这些检测器中的通道移动及其对分辨率降低的直接影响,提出了这些检测器最方便的测量条件。最后,建议在进行长时间测量之前,需要“等待时间”以避免信道移位效应。 (C)2016 Elsevier Ltd.保留所有权利。