Calibration of a 2D-CDB spectrometer using a reference 133Ba source

机译:使用参考133Ba光源校准2D-CDB光谱仪

获取原文
获取原文并翻译 | 示例
           

摘要

A procedure for calibrating a FAST-ComTec 2D-CDB spectrometer using a 133Ba source is presented.The energy calibration consisted of acquiring two-dimensional (2D) spectra by measuring simultaneously with two HPGe detectors the 356 and 511 keV peaks.As the 133Ba source only emits one photon per decay,the Ba contribution to the 2D spectra was built up by taking two successive and uncorrelated events,i.e.one from each detector (pseudo-coincidence technique).The FWHM of the spectrometer was estimated from the E1=E2 diagonal of the spectra.
机译:提出了使用133Ba光源校准FAST-ComTec 2D-CDB光谱仪的程序。能量校准包括通过同时使用两个HPGe检测器测量356和511 keV峰值来获取二维(2D)光谱,作为133Ba光源。每次衰变仅发射一个光子,通过采取两个连续且不相关的事件(即每个探测器发出的事件)(伪重合技术)建立了Ba对2D光谱的贡献。光谱