Three-dimensional dose distribution in contrast-enhanced digital mammography using Gafchromic XR-QA2 films: Feasibility study

机译:使用Gafchromic XR-QA2胶片的对比增强型乳腺X线摄影中的三维剂量分布:可行性研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

This study was aimed to establish three-dimensional dose distributions for contrast-enhanced digital mammography (CEDM) using self-developed Gafchromic XR-QA2 films. Dose calibration and distribution evaluations were performed on a full-field digital mammography unit with dual energy (DE) contrast-enhanced option. Strategy for dose calibration of films in the DE mode was based on the data obtained from common target/filter/kVp combinations used clinically and the dose response model modified from Rampado's model. Dose derived from films were also verified by measured data from an ionization chamber. The average difference of dose was 8.9% in the dose range for clinical uses. Three-dimensional dose distributions were estimated using triangular acrylic phantom equipped with the mammography system. Five pieces of film sheets were separately placed between the acrylic slabs to evaluate the dose distribution at different depths. After normalizing the dose in each pixel to the maximum dose at the top-center position of the acrylic, normalized dose distribution for transverse, coronal and sagittal planes, could thus be obtained. The depth dose distribution evaluated in this study may further serve as a reference for evaluating the patient glandular dose at different depths based on the entrance exposure information.
机译:这项研究的目的是使用自行开发的Gafchromic XR-QA2胶片,为对比增强型乳腺X线摄影(CEDM)建立三维剂量分布。剂量校准和分布评估是在具有双能(DE)增强对比功能的全场数字乳房X线照相设备上进行的。 DE模式下胶片剂量校准的策略是基于从临床上常用的靶标/滤光片/ kVp组合获得的数据,以及根据Rampado模型修改的剂量响应模型。还通过电离室的测量数据验证了源自薄膜的剂量。在临床使用剂量范围内,平均剂量差异为8.9%。使用配备了乳腺X线摄影系统的三角形丙烯酸模型来估计三维剂量分布。将五片薄膜片分别放置在丙烯酸板之间,以评估不同深度的剂量分布。在将每个像素的剂量归一化到丙烯酸的顶部中心位置处的最大剂量之后,可以因此获得横向,冠状和矢状面的归一化剂量分布。在这项研究中评估的深度剂量分布可进一步作为参考,基于入口暴露信息评估不同深度的患者腺体剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号