Eye lens dosimetry for interventional procedures - Relation between the absorbed dose to the lens and dose at measurement positions

机译:介入手术的眼镜片剂量法-眼镜片吸收剂量与测量位置剂量之间的关系

获取原文
获取原文并翻译 | 示例
       

摘要

This study investigated the relationship between the absorbed dose to the lens of the eye and the absorbed dose at different measurement positions near the eye of interventional radiologists. It also visualised the dose distribution inside the head, both when protective eyewear were used and without such protection. The best position for an eye lens dosimeter was found to be at the side of the head nearest to the radiation source, close to the eye. Positioning the dosimeter at the eyebrow could lead to an underestimation of the lens dose of as much as 45%. The measured dose distribution showed that the absorbed dose to the eye lenses was high compared to the other parts of the head, which stresses the importance of wearing protective eyewear. However, many models of eyewear were found to be deficient as the radiation could slip through at several places, e.g. at the cheek. The relationship between the absorbed dose to the lens and the kerma-area-product (P_(KA)) delivered to the patient was also studied.
机译:这项研究调查了眼部晶状体的吸收剂量与介入放射科医生眼睛附近不同测量位置的吸收剂量之间的关系。在使用防护眼镜和没有防护眼镜的情况下,它还可视化了头部内部的剂量分布。发现眼镜片剂量计的最佳位置是在靠近辐射源,靠近眼睛的头部侧面。将剂量计放在眉毛上可能会导致低估镜片剂量达45%。测得的剂量分布表明,与头部其他部位相比,对眼镜片的吸收剂量较高,这强调了戴防护眼镜的重要性。但是,发现许多型号的眼镜都存在缺陷,因为辐射可能会在几个地方(例如,在脸颊上。还研究了晶状体的吸收剂量与传递给患者的角膜面积积(P_(KA))之间的关系。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号