An analysis of energy deposition in a tissue equivalent proportional counter onboard the space shuttle

机译:航天飞机上组织当量比例计数器中的能量沉积分析

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

An improved prediction for space radiations in the lower earth orbits measured by the shuttle TEPC is obtained when energy loss straggling and chord length distribution of the detector are considered. A generalized analytic model is used to describe the energy deposition of direct ion interaction events in a micron-size detector. The transport calculation accounting for the shuttle configuration is accomplished by using a new version of HZETRN that has been extensively verified with laboratory and flight data. The agreement of predicted and measured lineal energy spectra is within 70% for the region above 2 keV/mu m but within a factor of 2.3 underpredicted for the region below this value. The inclusion of indirect delta ray events in the model is needed before possible causes for the underprediction below 2 keV/mu m can be assessed. (C) 1999 Elsevier Science Ltd. All rights reserved. [References: 42]
机译:当考虑到能量损失散布和探测器的弦长分布时,可以得到由航天飞机TEPC测量的对较低地球轨道空间辐射的改进预测。广义分析模型用于描述微米级检测器中直接离子相互作用事件的能量沉积。航天飞机配置的运输计算是通过使用HZETRN的新版本完成的,该版本已通过实验室和飞行数据进行了广泛验证。对于高于2 keV /μm的区域,预测和测量的线性能谱的一致性在70%以内,但对于低于该值的区域,预测的与测量的线性能谱相差2.3。在评估低于2 keV /μm的预测不足的可能原因之前,需要在模型中包含间接delta射线事件。 (C)1999 Elsevier ScienceLtd。保留所有权利。 [参考:42]