首页> 外文期刊>Radiation measurements >Performance of several active personal dosemeters in interventional radiology and cardiology
【24h】

Performance of several active personal dosemeters in interventional radiology and cardiology

机译:几种有源个人剂量计在介入放射学和心脏病学中的性能

获取原文
获取原文并翻译 | 示例
           

摘要

Active personal dosemeters (APDs) are very useful instruments for optimizing radiation protection of workers and for increasing worker's awareness of unexpected exposures. The challenge of monitoring personal equivalent doses with APDs in interventional fluoroscopy is that they must be sensitive to low energy photon beams and be able to record high dose rates. The aim of this work is to verify both the performance and the reliability of four active personal dosemeters (APDs) and one direct ion storage (DIS) dosemeter in typical X-ray radiation fields used during interventional radiology and cardiology (IR/IC) procedures. The values of the personal dose equivalent at a depth of 10 mm measured by the APDs are compared with the response of a whole body thermoluminescence dosemeter (TLD) as a reference dosemeter. The response is found to be satisfactory in the tested situations.
机译:主动式个人剂量计(APD)是非常有用的工具,可优化工作人员的辐射防护并提高工作人员对意外暴露的认识。在介入荧光透视法中用APD监测个人等效剂量的挑战在于,它们必须对低能光子束敏感并能够记录高剂量率。这项工作的目的是验证介入放射学和心脏病学(IR / IC)程序中使用的典型X射线辐射场中的四个有源个人剂量计(APD)和一个直接离子存储(DIS)剂量计的性能和可靠性。 。将通过APD测量的在10毫米深度处的个人剂量当量值与作为参考剂量计的全身热致发光剂量计(TLD)的响应进行比较。发现该响应在测试情况下令人满意。