首页> 外文期刊>Radiation measurements >Calibration of the Makrofol-DE nuclear track detector using relativistic lead ions
【24h】

Calibration of the Makrofol-DE nuclear track detector using relativistic lead ions

机译:使用相对论性铅离子对Makrofol-DE核径迹探测器进行校准

获取原文
获取原文并翻译 | 示例

摘要

We present the calibration of the Makrofol-DE nuclear track detector (NTD) using Pb-ions of 158 GeV/amu. Improvements in the post-etched surfaces, reduction in the surface background of the Makrofol-DE NTDs and high-contrast tracks were achieved with the appropriate addition of ethyl alcohol in KOH aqueous solutions. The calibration of Makrofol-DE has shown for the first time all the peaks arising due to nuclear fragments with Z >= 52. The measurement of the cone heights shows well-separated individual peaks for Z/beta = 59-82 and 83 (charge pickup). (c) 2005 Elsevier Ltd. All rights reserved.
机译:我们目前使用158 GeV / amu的Pb离子对Makrofol-DE核径迹检测器(NTD)进行校准。通过在KOH水溶液中适当添加乙醇,可以提高后蚀刻表面的质量,减少Makrofol-DE NTD的表面本底和提高对比度。 Makrofol-DE的校准首次显示了由于Z> = 52的核碎片而产生的所有峰。视锥高度的测量显示Z / beta = 59-82和83(电荷拾起)。 (c)2005 Elsevier Ltd.保留所有权利。