Dosimetry in mixed gamma-neutron radiation fields and energy compensation filters for CaF2 : Tm TL detectors

机译:CaF2:Tm TL探测器的混合γ中子辐射场的剂量测定法和能量补偿滤光片

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Strong fields of thermal and epithermal neutrons as used for Boron Neutron Capture Therapy (BNCT) show a relatively small Gamma dose. However, for accurate dose estimation, both the neutron as well as the gamma component has to be considered. Different approaches have been studied with various thermoluminescence detectors (TLDs). As a result, CaF2:Tm detectors (TLD-300) are attributed to be most insensitive to thermal neutrons but not tissue-equivalent. In this paper we focus on the measurement of the gamma dose with TLD-300 in mixed irradiation fields. To overcome the disadvantage of being not tissue-equivalent, a study on energy compensation filters was performed. The paper comprises also calibrations of TLD-300 using energy compensation filters at the PTB-reference field for thermal neutrons, and first test experiments at the TRIGA Mainz including simulations.
机译:硼中子俘获疗法(BNCT)中使用的热中子和超热中子的强场显示出相对较小的Gamma剂量。但是,为了进行准确的剂量估算,必须同时考虑中子和伽马分量。已经使用各种热致发光检测器(TLD)研究了不同的方法。结果,CaF2:Tm检测器(TLD-300)被认为对热中子最不敏感,但对组织等效不敏感。在本文中,我们着重于在混合辐射场中使用TLD-300测量伽马剂量。为了克服不等同于组织的缺点,对能量补偿滤波器进行了研究。本文还包括在热中子的PTB参考场上使用能量补偿滤波器对TLD-300进行的校准,以及在TRIGA Mainz进行的包括模拟在内的首次测试实验。