The effect of accurate positioning of TLD-100 samples in the center of the reader planchet on precision and kinetic parameters was investigated. Significant improvement in precision is obtained by careful positioning of the sample in the center of the planchet, by as much as 40%, 20% and 30% using different methods of glow curve analysis to estimate the intensity of the U signal. (C) 2014 Elsevier Ltd. All rights reserved.
机译:研究了将TLD-100样品准确定位在阅读器的棘轮中心对精度和动力学参数的影响。通过使用不同的辉光曲线分析方法估算U信号的强度,将样品小心地定位在棘轮的中心,可以将精度显着提高40%,20%和30%。 (C)2014 Elsevier Ltd.保留所有权利。