Energy spectra in water for the 6 MV X-ray therapeutic beam generated by Clinac-2300 linac

机译:Clinac-2300直线加速器产生的6 MV X射线治疗光束在水中的能谱

获取原文
获取原文并翻译 | 示例
           

摘要

There is a lack of extensive data comprising energy spectra of therapeutic beams used in tele-radiotherapy, generated by medical linear accelerators. In particular there is a lack of the data for energy spectra in water. However, the spectra in water differ from those in air significantly because of strong photon scattering processes. The aim of this paper was the presentation of the wide range of detailed data for the 6 MV X-ray therapeutic beams from a medical linear accelerator Clinac-2300 by Varian. The presented data were derived by the use of Monte Carlo computer simulations (GEANT4 code). The performed investigations indicate that shapes of the spectra, total number of photon registered in a bin as well as mean energy of the considered beam depend on a depth in water, on a distance from the central-axis of the beam and on a radiation field size. However, shapes of the spectra as well as the beam mean energy does not depend on a bin size, independently of a depth in water and a radiation field whereas total number of photon registered in a bin is related to a bin size. Majority of the presented results were obtained for a cylindrical 1.41 cm(3) bin with the radius of 1.5 cm. The obtained data are very useful for the accurate absorbed depth-dose determination particularly outside the reference conditions, for advanced treatment planning systems, for constructors of medical accelerators etc. (C) 2014 Elsevier Ltd. All rights reserved.
机译:缺乏包含由医学线性加速器产生的,用于远距离放射治疗的治疗束能谱的广泛数据。特别是缺乏水中能谱的数据。但是,由于强的光子散射过程,水中的光谱与空气中的光谱明显不同。本文的目的是介绍来自瓦里安(Varian)的医用线性加速器Clinac-2300的6 MV X射线治疗束的大量详细数据。呈现的数据是通过使用蒙特卡洛计算机模拟(GEANT4代码)得出的。进行的研究表明,光谱的形状,在一个仓中记录的光子总数以及所考虑的光束的平均能量取决于水中的深度,与光束中心轴的距离以及辐射场。尺寸。然而,光谱的形状以及束平均能量不依赖于箱的大小,而与水的深度和辐射场无关,而在箱中记录的光子总数与箱的大小有关。呈现的结果的大部分是获得的,圆柱体为1.41 cm(3),半径为1.5 cm。所获得的数据对于准确确定吸收的深度剂量非常有用,特别是在参考条件之外,对于先进的治疗计划系统,医疗加速器的制造商等。(C)2014 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号