Anodic stripping voltammetric determination of uranium at a thin palladium film-aluminum electrode: Analysis of some uranium mineral ores

机译:钯薄膜铝电极上阳极溶出伏安法测定铀:一些铀矿矿石的分析

获取原文
获取原文并翻译 | 示例
           

摘要

In the present work, a rapid deposition anodic stripping voltammetry (ASV) for determination of uranium is presented. For this purpose, the uranyl hexacyanoferrate (K2UO2[Fe(CN)(6)]) is deposited electrochemically on a thin palladium-aluminum electrode (Pd-Al) from a UO22+ solution in the presence of K3Fe(CN)(6). Then, the well stable (K2UO2[Fe(CN)(6)]) on the electrode was stripped by anodic differential pulse voltammetry for measuring the UO22+ ion concentration. The effect of operational parameters, including: concentration of K3Fe(CN)(6), solution pH, deposition potential, and deposition time were studied. In optimum conditions, the calibration graph was linear in the concentration range 10(-5)-7 x 10(-4) mol L-1 with a detection limit of 6.2 x 10(-6) mol The influence of some concomitant ions in K2UO2[Fe(CN)(6)] formation was investigated. The proposed method was used for the rapid determination of uranium in sonic uranium mineral ores.
机译:在目前的工作中,提出了一种用于沉积铀测定的快速沉积阳极溶出伏安法(ASV)。为此,在存在K3Fe(CN)(6)的情况下,从UO22 +溶液将六氰合铁酸铀酰(K2UO2 [Fe(CN)(6)])电化学沉积在薄钯铝电极(Pd-Al)上。然后,通过阳极差分脉冲伏安法剥离电极上的稳定性良好的(K2UO2 [Fe(CN)(6)]),以测量UO22 +离子浓度。研究了操作参数的影响,包括:K3Fe(CN)(6)的浓度,溶液的pH值,沉积电势和沉积时间。在最佳条件下,校准曲线在浓度范围10(-5)-7 x 10(-4)mol L-1内呈线性,检出限为6.2 x 10(-6)mol。研究了K2UO2 [Fe(CN)(6)]的形成。所提出的方法用于快速测定声铀矿物矿石中的铀。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号