Performance of the Packard Tri-Carb (R) 2770TR/SL liquid scintillation analyzer for C-14 dating

机译:Packard Tri-Carb(R)2770TR / SL液体闪烁分析仪针对C-14测年的性能

获取原文
获取原文并翻译 | 示例
           

摘要

We present results that demonstrate the potential of the Packard Tri-Carb(R) Model 2770TR/SL for radiocarbon dating. For 2 g of sample benzene, a stable background count rate of 0.307 cpm and a stable counting efficiency of 64.78% were determined using a 13-75 keV counting window. Changes to the mathematical routines for t-SIE (quench indicating parameter) calculation and a reduction in the activity of the external standard have enabled stability of the t-SIE to be achieved, and combined with the use of a suitable balance point counting window; all of these factors give the stability of performance required for C-14 dating. Calculations based on the above parameters indicate that the limit of detection for 2 g samples, counted for 5000 min, is >48,900 yr sp. The great advantage of this system is that these data were acquired using inexpensive standard 7-mL low potassium borosilicate glass vials. Vial holders manufactured from BGO reduced the background to 0.15 cpm with a minimum effect on efficiency (64.46% for 13-75 keV). A similar calculation of the limit of detection gave >51,700 sp. The use of the BGO vial holders in other instruments employing time-resolved liquid scintillation counting (TR-LSC) (Models 2250CA and 2260XL) also brought about significant improvements in detection limits. [References: 15]
机译:我们提供的结果证明了Packard Tri-Carb(R)2770TR / SL型放射性碳测年的潜力。对于2 g的样品苯,使用13-75 keV计数窗口确定了稳定的背景计数率0.307 cpm和稳定的计数效率64.78%。改变t-SIE(淬火指示参数)计算的数学程序并减少外标的活性,可以实现t-SIE的稳定性,并结合使用合适的平衡点计数窗口;所有这些因素都提供了C-14约会所需的性能稳定性。根据上述参数进行的计算表明,计数为5000分钟的2 g样品的检出限为> 48,900 yr sp。该系统的最大优势在于,这些数据是使用便宜的标准7毫升低硼硅酸钾玻璃小瓶采集的。由BGO制造的样品瓶架将背景降低到0.15 cpm,对效率的影响最小(13-75 keV时为64.46%)。相似的检测限计算得出> 51,700 sp。在其他使用时间分辨液体闪烁计数(TR-LSC)的仪器(型号2250CA和2260XL)中使用BGO样品瓶支架也可以显着提高检测限。 [参考:15]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号