Energy dependence of the supralinearity (f(D)max) of peaks 7 and 8 in the high temperature thermoluminescence of LiF:Mg,Ti (TLD-100): Interpretation using the unified interaction model

机译:LiF:Mg,Ti(TLD-100)高温热致发光中峰7和8的超线性(f(D)max)的能量依赖性:使用统一相互作用模型的解释

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

It is demonstrated that the supralinearity of the dose response of glow peaks 7, 8 in the glow curve of LiF:Mg,Ti (TLD-100) are very strongly dependent on photon/electron energy. Previously published data on f(D)_(max) at photon energies of 1.25 MeV, 100 keV and 8.1 keV effective energy coupled with new data at ~ 540 keV using ~(90)Sr/~(90)Y beta rays reveals that the maximum supralinearity f(D)_(max) decreases from values of ~200 and ~30 at 1.25 MeV, through intermediate values at 540 keV and 100 keV, to values of ~ 30 and ~3 at 8.1 keV effective energy. The normalized dose response f(D) for all energies is modeled using the Unified Interaction Model and the dependence of f(D)_(max) on energy is interpreted as arising from strong dependence of the relative intensity of localized recombination on particle energy (ionization density).
机译:结果表明,LiF:Mg,Ti(TLD-100)辉光曲线中辉光峰7、8剂量响应的超线性非常强烈地依赖于光子/电子能量。先前发表的有关光子能量为1.25 MeV,100 keV和8.1 keV的有效能量下f(D)_(max)的数据,再加上〜(90)Sr /〜(90)Yβ射线在〜540 keV处的新数据,揭示了最大超线性度f(D)_(max)从1.25 MeV时的〜200和〜30值,通过540 keV和100 keV的中间值,降至8.1 keV的有效能量的〜30和〜3值。使用统一相互作用模型对所有能量的归一化剂量响应f(D)进行建模,并将f(D)_(max)对能量的依赖性解释为是由于局部复合的相对强度对粒子能量的强烈依赖性引起的(电离密度)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号