Bland-Altman analysis is used to compare two different methods of measurement and to determine whether a new method of measurement may replace an existing accepted 'gold standard' method. In this work, Bland-Altman analysis has been applied to radiation dosimetry to compare the PTW Markus and Roos parallel plate ionisation chambers and a PTW PinPoint chamber against a Farmer type ionisation chamber which is accepted as the gold standard for radiation dosimetry in the clinic. Depth doses for low energy x-rays beams with energies of 50, 75 and 100 kVp were measured using each of the ionisation chambers. Depth doses were also calculated by interpolation of the data in the British Journal of Radiology (BJR) Report 25. From the Bland-Altman analysis, the mean dose difference between the two parallel plate chambers and the Farmer chambers was 1% over the range of depths measured. The PinPoint chamber gave significant dose differences compared to the Farmer chamber. There were also differences of up to 12% between the BJR Report 25 depth doses and the measured data. For the Bland-Altman plots, the lines representing the limits of agreement were selected to be a particular percentage agreement e.g. 1 or 2%, instead of being based on the standard deviation (σ) of the differences. The Bland-Altman statistical analysis is a powerful tool for making comparisons of ionisation chambers with an ionisation chamber that has been accepted as a 'gold standard'. Therefore we conclude that Bland-Altman analysis does have a role in assessing radiation dosimeter performance relative to an established standard.
机译:使用Bland-Altman分析来比较两种不同的测量方法,并确定一种新的测量方法是否可以替代现有的公认“金标准”方法。在这项工作中,将Bland-Altman分析应用于放射线剂量测定,以将PTW Markus和Roos平行板电离室和PTW PinPoint室与Farmer型电离室进行比较,后者被公认为是放射线剂量测定的金标准。使用每个电离室测量能量分别为50、75和100 kVp的低能X射线束的深度剂量。还通过对《英国放射学杂志》(BJR)报告25中的数据进行插值来计算深度剂量。根据Bland-Altman分析,两个平行板腔室和Farmer腔室之间的平均剂量差在1%的范围内。测量的深度。与农夫室相比,PinPoint室的剂量差异很大。 BJR报告25深度剂量与测量数据之间也存在高达12%的差异。对于布兰德-奥特曼(Bland-Altman)图,代表协议限制的线选择为特定百分比协议,例如1%或2%,而不是基于差异的标准偏差(σ)。 Bland-Altman统计分析是比较电离室和已被接受为“金标准”的电离室的强大工具。因此,我们得出的结论是,相对于既定标准,Bland-Altman分析确实在评估辐射剂量计性能方面具有作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号