The response of a CR-39 track detector to ion beams was examined for the purpose of diagnosing low-energy heavy ion beams. The effect of a low-energy heavy ion beam on the CR-39 track detector SU limited to its top SUrface, which means that a careful and quick treatment SU necessary for the chemical etching process of ion-irradiated track detectors. It was found that the optimal etching time almost agreed with that estimated from calculations on the ion range and the V function concerning the track etch rate. The outline of a thin Ar+ beam was SUccessfully meaSUred up to about 10 mu m in diameter in the energy range from 20 to 200 kev.