Real-time estimation of dose reduction for pediatric CT using bismuth shielding (Conference Paper)

机译:使用铋屏蔽实时估算儿科CT剂量的减少量(会议论文)

获取原文
获取原文并翻译 | 示例

摘要

The purpose of this study is to estimate the dose reduction of bismuth shielding for pediatric CT using a real-time CT dose probe. Routine CT examinations of head, chest and abdomen were performed on a 10-year pediatric anthropomorphic phantom, with and without bismuth shield. A solid-state dosimeter was used to obtain real-time dose measurements and to estimate the dose reduction. The maximum dose reduction of eyes using bismuth shield for in-plane CT scanning can achieve 56% during spiral head scanning interval. The average dose reduction is around 25% for in-plane CT scanning, and reaches 62% combining the technique of automatic tube current modulation. The solid-state probe can provide fast and convenient measurements for dose estimation of superficial organs. It is suggested to use Bismuth shielding in routine CT examinations to protect the radiosensitive tissues.
机译:这项研究的目的是使用实时CT剂量探针来评估小儿CT铋屏蔽的剂量降低。头,胸和腹部的常规CT检查是在有或没有铋保护罩的10年期儿童拟人化体模上进行的。固态剂量计用于获得实时剂量测量并估算剂量减少。使用铋罩进行平面CT扫描时,眼睛的最大剂量减少可以在螺旋头扫描间隔内达到56%。平面CT扫描的平均剂量减少量约为25%,结合自动管电流调制技术可达到62%。固态探针可以为表浅器官的剂量估计提供快速方便的测量。建议在常规CT检查中使用铋屏蔽,以保护放射敏感性组织。