Best practices concerning qualification tests of radon passive detectors recommend the application of radon standard atmospheres at controlled conditions. A measurement technique has been developed, based on the application of the new on/off alpha track passive radon detector developed at ENEA. The capability of switch and pumping functions of this device has removed the time lag due to the radon transfer inside its sensitive volume and the post-exposure due to the residual radon decay. A special exposure facility has been developed based on a set of the above passive detectors connected with flange adapters to a radon chamber wall. This apparatus can be used substantially as a radon exposure standard (RES). Results of standardization tests at the ENEA Radon Facilities are given. The paper also addresses requirements for simple and small exposure facilities that permit accurate timed exposures and can be used for testing passive devices. (c) 2005 Elsevier Ltd. All rights reserved.
机译:有关ra无源探测器鉴定测试的最佳实践建议在受控条件下使用ra标准大气。基于在ENEA开发的新型开/关alpha轨道无源ra检测器的应用,已经开发了一种测量技术。该设备的开关和泵浦功能已消除了由于sensitive在其敏感体积内转移而造成的时间滞后以及由于残留ra衰变而导致的后暴露。基于一组上述无源探测器,并通过法兰适配器将其连接到a室壁上,已经开发了一种特殊的曝光设备。该装置基本上可以用作a暴露标准(RES)。给出了在ENEA Radon工厂的标准化测试结果。本文还介绍了对简单和小型曝光设备的要求,这些设备允许精确的定时曝光并可以用于测试无源设备。 (c)2005 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号