首页> 外文期刊>Radiation Physics and Chemistry >Evaluation of radiation dose to neonates in a special care baby unit
【24h】

Evaluation of radiation dose to neonates in a special care baby unit

机译:评估特殊护理婴儿病房对新生儿的辐射剂量

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The purpose of this study was to evaluate the patient entrance surface dose (ESD), organ dose and effective dose for neonates in the special care baby unit (SCBU) up to 28 days after birth. A total of 135 patients were examined during 4 months. ESDs were calculated from patient exposure parameters using DosCal software. Effective doses were calculated using software from the National Radiological Protection Board (NRPB). The mean patient ESD per procedure was 80 ± 0.02 μGy. The mean and range ; of the effective dose per procedure were 0.02 (0.01-0.3) mSv. The radiation dose in this study was higher compared to previous studies. A dedicated X-ray machine with additional filtration is recommended for patient dose reductions.
机译:这项研究的目的是评估出生后28天以内的特殊护理婴儿病房(SCBU)的新生儿患者的入口表面剂量(ESD),器官剂量和有效剂量。在4个月内检查了135位患者。使用DosCal软件根据患者的暴露参数计算ESD。使用国家放射防护局(NRPB)的软件计算有效剂量。每个程序的平均患者ESD为80±0.02μGy。均值和范围;每个程序的有效剂量为0.02(0.01-0.3)mSv。与以前的研究相比,该研究中的辐射剂量更高。建议使用带有额外过滤功能的专用X射线机以减少患者剂量。