Study of the spatial distribution of the absorbed dose in blood volumes irradiated using a teletherapy unit

机译:使用远程治疗仪辐射的血容量中吸收剂量的空间分布研究

获取原文
获取原文并翻译 | 示例
       

摘要

Blood irradiation can be performed using a dedicated blood irradiator or a teletherapy unit. A thermal device providing appropriate storage conditions during blood components irradiation with a teletherapy unit has been recently proposed. However, the most appropriated volume of the thermal device was not indicated. The goal of this study was to indicate the most appropriated blood volume for irradiation using a teletherapy unit in order to minimize both the dose heterogeneity in the volume and the blood irradiation time using these equipments. Theoretical and experimental methods were used to study the dose distribution in the blood volume irradiated using a linear accelerator and a cobalt-60 therapy machine. The calculation of absorbed doses in the middle plane of cylindrical acrylic volumes was accomplished by a treatment planning system. Experimentally, we also used cylindrical acrylic phantoms and thermoluminescent dosimeters to confirm the calculated doses. The data obtained were represented by isodose curves. We observed that an irradiation volume should have a height of 28 cm and a diameter of 28 cm and a height of 35 cm and a diameter of 35 cm, when the irradiation is to be performed by a linear accelerator and a cobalt-60 teletherapy unit, respectively. Calculated values of relative doses varied from 93% to 100% in the smaller volume, and from 66% to 100% in the largest one. A difference of 5.0%, approximately, was observed between calculated and experimental data. The size of these volumes permits the irradiation of blood bags in only one bath without compromising the homogeneity of the absorbed dose over the irradiated volume. Thus, these irradiation volumes can be recommend to minimize the irradiation time when a teletherapy unit is used to irradiate blood.
机译:可以使用专用血液辐照器或远距治疗仪进行血液辐照。最近已经提出了一种热装置,该热装置在用远程治疗单元照射血液成分期间提供适当的存储条件。然而,没有指出热装置的最合适的体积。这项研究的目的是指出最适合使用远程治疗设备进行照射的血液量,以便使使用这些设备的剂量异质性和血液照射时间最小化。理论和实验方法用于研究使用线性加速器和Cobalt-60治疗机照射的血量的剂量分布。通过处理计划系统可以计算圆柱状丙烯酸体积的中间平面中的吸收剂量。在实验上,我们还使用了圆柱形丙烯酸体模和热发光剂量计来确认计算出的剂量。所获得的数据由等剂量曲线表示。我们观察到,当要使用线性加速器和60钴远距放射治疗仪进行照射时,照射体的高度应为28 cm,直径为28 cm,高度为35 cm,直径为35 cm。 , 分别。相对剂量的计算值在较小体积中为93%至100%,在最大体积中为66%至100%。在计算和实验数据之间观察到大约5.0%的差异。这些体积的大小允许仅在一个浴中辐照血袋,而不会损害在辐照体积上吸收剂量的均匀性。因此,当使用远程治疗仪辐射血液时,建议使用这些辐射量以最大程度地减少辐射时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号