Agreement Between Omron 306 and Biospace InBody 720 Bioelectrical Impedance Analyzers (BIA) in Children and Adolescents

机译:欧姆龙306与Biospace InBody 720儿童和青少年生物电阻抗分析仪(BIA)之间的协议

获取原文
获取原文并翻译 | 示例

摘要

Purpose: The purpose of this study was to test the convergent validity of Omron 306 using Biospace InBody 720. Method: A total of 267 participants (145 boys; aged 10.4-17.9 years) completed testing during a single session. Each measure provided percent body fat (%BF), while the InBody 720 included fat-free mass (FFM). The validity was examined using the Pearson correlation. Limits of agreement (LOA) and multiple linear regression were also used to determine the impact of both age and FFM on the associations between the 2 measures. Results: The 2 measures of %BF were correlated by .63 (p<.001) in boys and .89 (p<.001) in girls. The mean difference (i.e., InBody - Omron) for %BF in boys was -4.7% with a lower LOA of -20.5% and upper LOA of 11.2%. The same comparison for girls resulted in a difference of 3.0% with a lower LOA of -10% and upper LOA of 4%. Examination of the residuals obtained from multiple linear regression indicated that FFM was the only statistically significant predictor of differences in boys (beta(FFM)=-0.25 +/- 0.08%, p=.001). There were no significant associations for girls. Conclusions: Findings indicate that estimates of %BF obtained from Omron in boys exceed estimates from InBody 720. Disagreement was evident in younger boys with lower levels of FFM. Girls' %BF was closer between the 2 bioelectrical impedance analyzer measures (less residual) with age and FFM not explaining the disagreement. Overall, the 2 measures were not equivalent.
机译:目的:本研究的目的是使用Biospace InBody 720测试Omron 306的收敛性有效性。方法:总共267名参与者(145名男孩; 10.4-17.9岁)在一次会议中完成了测试。每种方法都提供了身体脂肪百分比(%BF),而InBody 720则包含了无脂肪质量(FFM)。使用Pearson相关性检查有效性。一致性极限(LOA)和多元线性回归也用于确定年龄和FFM对这两种测量之间的关联的影响。结果:男孩的两项测量值%BF分别为0.63(p <.001)和女孩的0.89(p <.001)。男孩的%BF的平均差异(即InBody-Omron)为-4.7%,较低的LOA为-20.5%,较高的LOA为11.2%。对女孩的相同比较导致差异为3.0%,较低的LOA为-10%,较高的LOA为4%。从多元线性回归获得的残差的检验表明,FFM是男孩差异的唯一统计学上显着的预测因子(β(FFM)=-0.25 +/- 0.08%,p = .001)。没有明显的女孩协会。结论:研究结果表明,从欧姆龙获得的男孩%BF的估计值超过了InBody 720的估计值。FFM水平较低的年轻男孩显然存在分歧。女孩的%BF在两种生物电阻抗分析仪测量值(残差较少)与年龄之间的关系更接近,而FFM不能解释这一分歧。总体而言,这两项措施并不等同。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号