首页> 外文期刊>Radiography >Options for radiation dose optimisation in pelvic digital radiography: A phantom study
【24h】

Options for radiation dose optimisation in pelvic digital radiography: A phantom study

机译:骨盆数字X线摄影中辐射剂量优化的选项:幻像研究

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To investigate the effects of phantom orientation and AEC chamber selection on radiation dose and image quality (IQ.) for digital radiography (DR) examinations of the pelvis.Methods: A phantom study was conducted using a DR detector, utilising all AEC chamber combinations. Current recommended orientation (Cr-AEC) was with the outer AEC chambers cranially orientated. mAs (given), source-to-skin distance and kVp data facilitated entrance surface dose and effective dose calculations. Six anatomical areas were blindly graded by two observers (3-point scale) for IQ, Statistical differences in radiation dose were determined using the paired Student's t-test. IQdata was analysed for inter-observer variability (ICC) and statistical differences (Wilcoxon test).Results: Switching phantom orientation (caudally orientated outer AEC chambers: Ca-AEC) reduced mean radiation dose by 36.8%, (p < 0.001). A minor reduction in median IQ. (15.5 vs. 15) was seen (p < 0.001). One Ca-AEC orientated image (1.6%) had all anatomical areas graded inadequate' by at least one observer; all other images were considered 'adequate' for all areas. In the Ca-AEC orientation, at least a 44% dose reduction was achievable (p < 0.001) when only the outer AEC chambers were used. In the Cr-AEC orientation, at least 11% dose reduction was achieved (p < 0.001); here the central chamber was used alone, or in combination. IQ. scores fell, but remained 'adequate'.Conclusion: Switching pelvic orientation relative to AEC chamber position can optimise radiation dose during pelvic radiography. AEC chamber position should be clearly marked on equipment to facilitate this. AEC selection should be an active process.
机译:目的:研究体模方向和AEC腔室选择对骨盆数字X线检查(DR)的辐射剂量和图像质量(IQ。)的影响。方法:使用DR检测器进行体模研究,利用所有AEC腔室组合。当前推荐的方向(Cr-AEC)是外部AEC腔为颅骨方向。 mAs(给定),源到皮肤距离和kVp数据有助于入射表面剂量和有效剂量计算。由两名观察员(3分制)盲目对六个解剖区域进行智商分级,并使用配对的Student t检验确定辐射剂量的统计学差异。分析IQQdata的观察者间差异(ICC)和统计学差异(Wilcoxon检验)。中位数智商略有下降。观察到(15.5 vs.15)(p <0.001)。一张由Ca-AEC定向的图像(占1.6%)的所有解剖区域均被至少一名观察者分级为“不足”;所有其他图像都被认为在所有区域都“足够”。在Ca-AEC方向上,仅使用外部AEC腔室时,至少可以减少44%的剂量(p <0.001)。在Cr-AEC取向下,剂量减少至少11%(p <0.001);在这里,中央室可以单独使用,也可以组合使用。智商结论:改变盆腔相对于AEC腔位置的方向可以优化盆腔X线摄影的放射剂量。 AEC腔室的位置应在设备上清楚标出,以方便进行此操作。选择AEC应该是一个积极的过程。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号