Radiographic grids are used to improve image quality by absorbing scattered radiation before it reaches the image receptor. Grids are constructed of very thin strips of lead interspaced with aluminum. Using grids during portable examinations is common, especially when using digital imaging, such as computed radiography (CR) systems. A surgeon ordered a portable abdomen radiograph in the operating room following a lengthy abdominal surgery on a 56-year-old man. A CR cassette with a slip-on grid was used to obtain the image. The processed image reveals multiple, random linear densities (see Figure l). Upon further investigation, it was discovered that the grid was damaged, presumably from being dropped (see Figure 2).
机译:射线照相栅格用于通过在散射辐射到达图像接收器之前吸收其散射来改善图像质量。栅格由与铝隔开的非常细的铅条构成。在便携式检查期间通常使用栅格,尤其是在使用数字成像(例如计算机射线照相(CR)系统)时。一名外科医生在对一名56岁男子进行了长时间的腹部手术后,在手术室订购了一部便携式腹部X光片。使用带有滑入式栅格的CR暗盒获取图像。处理后的图像显示出多个随机的线性密度(见图1)。经过进一步调查,发现网格损坏了,大概是因为掉落了(参见图2)。