MR-Intracranial pressure (ICP): a method to measure intracranial elastance and pressure noninvasively by means of MR imaging: baboon and human study.

机译:MR颅内压(ICP):一种通过MR成像无创地测量颅内弹性和压力的方法:狒狒和人体研究。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To develop a noninvasive method for intracranial elastance and intracranial pressure (ICP) measurement. MATERIALS AND METHODS: Intracranial volume and pressure changes were calculated from magnetic resonance (MR) imaging measurements of cerebrospinal fluid (CSF) and blood flow. The volume change was calculated from the net transcranial CSF and blood volumetric flow rates. The change in pressure was derived from the change in the CSF pressure gradient calculated from CSF velocity. An elastance index was derived from the ratio of pressure to volume change. The reproducibility of the elastance index measurement was established from four to five measurements in five healthy volunteers. The elastance index was measured and compared with invasive ICP measurements in five patients with an intraventricular catheter at MR imaging. False-positive and false-negative rates were established by using 25 measurements in eight healthy volunteers and six in four patients with chronically elevated ICP. R:ESULTS: The mean of the fractional SD of the elastance index in humans was 19.6%. The elastance index in the five patients with intraventricular catheters correlated well with the invasively measured ICP (R:(2) = 0.965; P: <.005). MR imaging-derived ICPs in the eight healthy volunteers were 4.2-12.4 mm Hg, all within normal range. Measurements in three of the four patients with chronically elevated ICP were 20.5-34.0 mm Hg, substantially higher than the normal limit. CONCLUSION: MR imaging-derived elastance index correlates with ICP over a wide range of ICP values. The sensitivity of the technique allows differentiation between normal and elevated ICP.
机译:目的:开发一种颅内弹性和颅内压(ICP)测量的非侵入性方法。材料与方法:根据脑脊液(CSF)和血流的磁共振(MR)成像测量结果计算颅内体积和压力变化。由净经颅脑脊液和血液体积流速计算体积变化。压力变化是从CSF速度计算得出的CSF压力梯度变化中得出的。弹性指数是由压力与体积变化之比得出的。弹性指数测量的可重复性是在五名健康志愿者中从四次测量到五次测量确定的。测量了弹性指数,并将其与5例有MR成像的心室内导管的患者的有创ICP测量进行了比较。通过对8位健康志愿者和25位ICP慢性升高的患者中的6位进行25次测量,建立了假阳性和假阴性率。 R:结果:人体弹性指数的标准差平均值为19.6%。 5例脑室内导管患者的弹性指数与有创测量的ICP密切相关(R:(2)= 0.965; P:<.005)。八名健康志愿者的MR成像ICPs为4.2-12.4 mm Hg,均在正常范围内。 ICP慢性升高的四位患者中有三位的测量值为20.5-34.0 mm Hg,大大高于正常限值。结论:在很大的ICP值范围内,MR成像的弹性指数与ICP相关。该技术的灵敏度可以区分正常ICP和升高的ICP。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号