首页> 外文期刊>Radiology >In-line pressures generated in small-bore central venous catheters during power injection of CT contrast media.
【24h】

In-line pressures generated in small-bore central venous catheters during power injection of CT contrast media.

机译:在CT造影剂动力注入过程中,小口径中心静脉导管中产生的管路压力。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To determine in-line pressures generated in small-bore central venous catheters during power injection of computed tomographic (CT) contrast media. MATERIALS AND METHODS: Five 3.0-7.0-F central venous catheters for pediatric patients were tested at full and half lengths in vitro. In-line pressures were measured during power injection of three contrast media. Rates of injection were increased in steps from 0.1 to 5.0 mL/sec or until a peak pressure of 100 psi (700 kPa) was achieved. The maximum tolerated flow rate was determined with reference to the manufacturer's suggested operating pressure limit for each catheter. RESULTS: At full length, the maximum tolerated flow rates were as follows: 2-3 mL/sec for the large lumen and 1-1.4 mL/sec for the small lumen of the 7.0-F double-lumen catheter; 0.2-0.4 and 0.8-1.2 mL/sec for the 3.0- and 4.0-F peripherally inserted central catheters, respectively; 0.7-1.2 mL/sec for the 6.6-F catheter; and only 0.2 mL/sec for the 4.2-F catheter, which ruptured during testing at higher flow rates. CONCLUSION: Flow rates were documented at which certain small-bore central venous catheters should tolerate power injection of CT contrast media with peak pressures remaining below the manufacturer's recommended operating pressure limits. These data may serve as a guide for clinical use.
机译:目的:确定在计算机断层扫描(CT)造影剂动力注入过程中,小口径中心静脉导管中产生的管路压力。材料与方法:在体外测试了五根3.0-7.0-F小儿中心静脉导管。在三种造影剂的动力注入过程中测量在线压力。进样速度从0.1毫升/秒逐步增加到5.0毫升/秒,或者直到达到100 psi(700 kPa)的峰值压力。参考每个导管的制造商建议的操作压力极限,确定最大耐受流速。结果:全长时,最大耐受流速如下:7.0-F双腔导管的大内腔为2-3 mL / sec,小腔为1-1.4 mL / sec;对于3.0-F和4.0-F外围插入的中央导管,分别为0.2-0.4和0.8-1.2 mL / sec; 6.6-F导管为0.7-1.2 mL /秒; 4.2-F导管仅0.2 mL / sec,在测试过程中以更高的流速破裂。结论:记录了某些小口径中心静脉导管应耐受CT造影剂动力注入的流速,峰值压力仍低于制造商建议的操作压力极限。这些数据可以作为临床使用的指南。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号