首页> 外文期刊>Lymphology >Local tissue water changes assessed by tissue dielectric constant: single measurements versus averaging of multiple measurements.
【24h】

Local tissue water changes assessed by tissue dielectric constant: single measurements versus averaging of multiple measurements.

机译:通过组织介电常数评估局部组织水分变化:单次测量与多次测量的平均值。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Previous reports describe the use of average tissue dielectric constant (TDC) measurements to assess local tissue water and its change. Our goal was to determine if a single TDC measurement could be used in place of the average of multiple measurements. The comparison criteria used to test this was the extent to which single and averaged measurements yielded similar TDC values in both normal and lymphedematous tissue. Measurements were made on both ventral forearms to a depth of 2.5 mm in 10 women with unilateral arm lymphedema. The main results showed that the 95% confidence interval for differences between single and averaged TDC values was less than +/- 1 TDC unit for both normal and lymphedematous arms. This finding strongly suggests that for most, if not all, clinical evaluations, suitable assessments can be made using a single TDC measurement.
机译:以前的报告描述了使用平均组织介电常数(TDC)测量来评估局部组织水及其变化。我们的目标是确定是否可以使用单个TDC测量代替多个测量的平均值。用于测试此结果的比较标准是在正常和淋巴水肿组织中单次测量和平均测量得出相似的TDC值的程度。在10位单侧手臂淋巴水肿的女性中,对两个前臂的前臂进行了测量,测量深度为2.5 mm。主要结果表明,正常和淋巴水肿组的TDC值与平均TDC值之间的差异的95%置信区间均小于+/- 1 TDC单位。这一发现强烈表明,对于大多数(如果不是全部)临床评估,可以使用单个TDC测量进行适当的评估。