首页> 外文期刊>Langenbeck's archives of surgery >Which parameters are needed for targeting a multitined radiofrequency device--an approach to a simple algorithm.
【24h】

Which parameters are needed for targeting a multitined radiofrequency device--an approach to a simple algorithm.

机译:针对多种射频设备需要哪些参数-一种简单算法的方法。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND: The use of radiofrequency ablation (RFA) for treatment of liver malignancies is limited by the high rate of local recurrences. The aim of this experimental study was to evaluate parameters describing the reproducible target volume of a RFA procedure in order to facilitate better applicator placement. MATERIALS AND METHODS: RFA was performed in perfused and nonperfused pig livers. The following parameters were measured: axial and transverse diameter, front margin, coagulation center, diameter of sphere ablated (D(S)), distance to center (DC), and volume. Graphic overlays were utilized to visualize variability. Parameters were evaluated for Rita XL (2 algorithms), LeVeen, and Rita Xli applicators. RESULTS: The best prediction of a reproducibly ablated target volume can be made by the diameter of the sphere ablated and the distance of the applicator tip to center of the sphere (DC). The spheres were significantly different in diameter (D(S)) depending on the applicator Rita XL 29 +/- 6 mm, Rita XL(wet) 35 +/- 5 mm, LeVeen 35 +/- 8 mm, Rita Xli 44 +/- 5 mm (perfused livers, p < 0.001). Graphic overlay demonstrated differences in variability that can influence the reliability of the system. CONCLUSIONS: D(S) and DC as specific values for each applicator and algorithm facilitate a placement of the applicator relative to the target volume that maximizes the chance of complete ablation.
机译:背景:射频消融(RFA)用于治疗肝恶性肿瘤的局限性在于局部复发率高。这项实验研究的目的是评估描述RFA程序可重现目标体积的参数,以便于更好地放置施药器。材料与方法:RFA在灌注和未灌注的猪肝中进行。测量了以下参数:轴向和横向直径,前缘,凝结中心,消融球的直径(D(S)),到中心的距离(DC)和体积。图形叠加被用来可视化可变性。对Rita XL(2种算法),LeVeen和Rita Xli涂药器的参数进行了评估。结果:可消融球体的直径和涂药器尖端到球体中心(DC)的距离可以对可重复消融的目标体积做出最佳预测。球体的直径(D(S))明显不同,具体取决于施涂器Rita XL 29 +/- 6 mm,Rita XL(湿)35 +/- 5 mm,LeVeen 35 +/- 8 mm,Rita Xli 44 + /-5毫米(肝脏灌注,p <0.001)。图形叠加图显示了可变性上的差异,这些差异会影响系统的可靠性。结论:D(S)和DC作为每个涂抹器和算法的特定值,有助于相对于目标体积放置涂抹器,从而最大程度地增加了完全消融的机会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号