Indirect and direct simultaneous, comparative blood pressure measurements with the Bosotron 2 device.

机译:Bosotron 2设备可进行间接和直接同时比较血压测量。

获取原文
获取原文并翻译 | 示例
           

摘要

Testing the accuracy of the automatic blood pressure measuring device Bosotron 2, we performed simultaneous, same-arm, comparative blood pressure measurements with three instruments (A, B, and C) of the device and by auscultation in the antecubital fossa by a trained observer in 24 volunteers each. The volunteers were evenly matched for age, sex, and blood pressure level. The mean systolic differences and their standard deviations (24 subjects each, n = 288 single measurements) were -1.32+/-3.8 mm Hg for instrument A, -0.99+/-3.9 mm Hg for instrument B, and -1.12+/-5.0 mm Hg for instrument C. The corresponding values for the diastolic differences were -2.71+/-4.4, -3.46+/-4.3, and -2.72+/-3.9 mm Hg, respectively. Combining the results for the three instruments, the comparison yielded a good accuracy (mean difference) and sufficient repeatability of the differences (standard deviation) for the systolic and diastolic blood pressures (systolic -1.14+/-4.3 mm Hg; diastolic -2.96+/-4.2 mm Hg). In addition to the British Hypertension Society protocol, in 33 patients with coronary artery disease the measurements using the three Bosotron 2 devices were compared with measurements within the aortic arch. The mean systolic/diastolic differences (+/-SD) were +1.45+/-4.3 and +7. 27+/-4.7 mm Hg for instrument A, +4.70+/-12.8 and +1.73+/-7.4 mm Hg for instrument B, and -3.74+/-8.0 and +5.61+/-2.6 mm Hg for instrument C. Combining the results of the three instruments, the blood pressure was determined to be only slightly higher (systolic +1.74+/-9.6 mm Hg; diastolic +4.87+/-5.6 mm Hg) by the Bosotron 2 R device as compared with the aortic arch pressure. The Bosotron 2 R device seems to be suitable for clinical use and for monitoring blood pressure during clinical-pharmacological studies.
机译:为了测试自动血压测量设备Bosotron 2的准确性,我们使用该设备的三种仪器(A,B和C)并在受过训练的观察者的前肘窝进行听诊的同时,进行了同一臂的比较血压测量每位24名志愿者。志愿者的年龄,性别和血压水平均相匹配。仪器的平均收缩压差及其标准偏差(每组24位受试者,n = 288次单次测量)对于器械A为-1.32 +/- 3.8 mm Hg,对于器械B为-0.99 +/- 3.9 mm Hg,和-1.12 +/-器械C的压力为5.0毫米汞柱。舒张压差异的相应值分别为-2.71 +/- 4.4,-3.46 +/- 4.3和-2.72 +/- 3.9毫米汞柱。结合这三种仪器的结果,比较得出的收缩压和舒张压(收缩压-1.14 +/- 4.3 mm Hg;舒张压-2.96+)具有良好的准确性(均值差)和足够的重复性可重复性(标准差)。 /-4.2毫米汞柱)。除了英国高血压学会的协议外,还将33例冠状动脉疾病患者中使用三种Bosotron 2装置的测量结果与主动脉弓内的测量结果进行了比较。平均收缩压/舒张压差(+/- SD)为+1.45 +/- 4.3和+7。仪器A为27 +/- 4.7 mm Hg,仪器B为+4.70 +/- 12.8和+1.73 +/- 7.4 mm Hg,仪器C为-3.74 +/- 8.0和+5.61 +/- 2.6 mm Hg。结合这三种仪器的结果,与主动脉相比,Bosotron 2 R装置确定的血压仅稍高(收缩压+1.74 +/- 9.6 mm Hg;舒张压+4.87 +/- 5.6 mm Hg)足弓压力。 Bosotron 2 R设备似乎适合于临床使用,并且适合在临床药理学研究期间监测血压。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号