首页> 外文期刊>Cardiovascular engineering >Construction of an Artificial Heart Pump Performance Test System
【24h】

Construction of an Artificial Heart Pump Performance Test System

机译:人工心脏泵性能测试系统的构建

获取原文
获取原文并翻译 | 示例
           

摘要

A hydraulic loop, which simulates pressure/ flow response of the human circulatory system, is needed to bench test the various versions of rotary left ventricular assist devices (LVADs). This article describes the design of such a loop and the simulated response of different physiological states, such as a healthy person in sleep, rest, and mild physical activity, and in different pathological states. The loop consists of: (1) pulsatile left and right cardiac simulators; (2) air/ water tanks to model the venous and arterial compliances; (3) tygon tubes to model the venous, arterial, and other system flow resistances; and (4) a tuning clamp to model the variation in system resistance characteristics under different cardiac pressure/flow conditions. The simulated responses were compared to the data found in the literature to validate the loop performance prior to LVAD testing.
机译:需要一个液压回路来模拟人体循环系统的压力/流量响应,以便对各种形式的左旋旋转辅助设备(LVAD)进行基准测试。本文介绍了这种循环的设计以及不同生理状态(例如健康的人在睡眠,休息和轻度的体育活动中以及处于不同病理状态)的模拟响应。该循环包括:(1)脉动的左右心脏模拟器; (2)空气/水箱,以模拟静脉和动脉的顺应性; (3)tygon管以模拟静脉,动脉和其他系统的流阻; (4)调谐钳,用于模拟在不同心压/流量条件下系统阻力特性的变化。将模拟响应与文献中的数据进行比较,以在LVAD测试之前验证环路性能。