Validation of intensity modulated radiation therapy patient plans with portal images

机译:用门户图像验证强度调制放射治疗患者计划

获取原文
获取原文并翻译 | 示例
       

摘要

The goal of this study was to show the feasibility of step and shoot intensity-modulated radiation therapy pre-treatment quality control for patients using the electronic portal imaging device (iViewGT) fitted on a Sli+ linac (Elekta Oncology Systems, Crawley, UK) instead of radiographic films. Since the beginning of intensity-modulated radiation therapy treatments, the dosimetric quality control necessary before treating each new patient has been a time-consuming and therefore costly obligation. In order to fully develop this technique, it seems absolutely essential to reduce the cost of these controls, especially the linac time. Up to now, verification of the relative dosimetry field by field has been achieved by acquiring radiographic films in the isocenter plane and comparing them to the results of the XiO planning system (Computerized Medical Systems, Missouri, USA) using RIT113 v4.1 software (Radiological Imaging Technology, Colorado, USA). A qualitative and quantitative evaluation was realised for every field of every patient. A quick and simple procedure was put into place to be able to make the same verifications using portal images. This new technique is not a modification of the overall methodology of analysis. The results achieved by comparing the measurement with the electronic portal imaging device and the calculation with the treatment planning system were in line with those achieved with the films for all indicators we studied (isodoses, horizontal and vertical dose profiles and gamma index).
机译:这项研究的目的是展示使用安装在Sli +直线加速器(Elekta Oncology Systems,Crawley,英国)上的电子门禁成像设备(iViewGT)对患者进行步进和发射强度调制的放射治疗预处理质量控制的可行性。胶卷胶片。自开始进行强度调制放射治疗以来,在治疗每位新患者之前必须进行剂量学质量控制是一项耗时且因此昂贵的义务。为了充分开发这种技术,减少这些控制装置的成本,尤其是直线加速器的时间,似乎绝对必要。到目前为止,通过在等角点平面上获取射线照相胶片并将其与XiO规划系统(美国密苏里州的计算机医疗系统)的结果进行比较,已经通过使用RIT113 v4.1软件(放射成像技术,美国科罗拉多州)。对每个患者的每个领域都进行了定性和定量评估。建立了一个快速简单的过程,以便能够使用门户网站映像进行相同的验证。这项新技术不是对整个分析方法的修改。通过比较电子门静脉成像设备的测量值和治疗计划系统的计算结果,与我们研究的所有指标(等剂量,水平和垂直剂量分布以及伽玛指数)的胶片所取得的结果一致。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号