INTRODUCTION: Verification of absorbed dose in target volume is a key factor for quality assurance in radiotherapy. In vivo measurements allow evaluation of the variations in dose with time and variations between measured doses and calculated doses by TPS. The aim of this work were to evaluate reproducibility of patient positioning and to compare calculated doses by 2 different TPS. PATIENTS AND METHODS: Twenty patients were divided in 2 groups according to the thickness of their breast (mean SSD = 92.9 cm). In vivo measurement was performed within the first two sessions. RESULTS: Reproducibility of SSD evaluation was made on 12 beams between 2 fractions. With a tolerance margin of 0.5 cm, positioning errors were present in 33% (4/12). The 2 TPS were in agreement in 75% (30/40). CONCLUSION: In vivo dosimetry can be a very interesting tool to assess patients positioning variations and TPS dose calculation.
机译:简介:验证目标剂量中的吸收剂量是放射治疗质量保证的关键因素。体内测量允许通过TPS评估剂量随时间的变化以及测量剂量和计算剂量之间的变化。这项工作的目的是评估患者定位的可重复性,并比较2种不同TPS的计算剂量。患者与方法:20例患者根据乳房的厚度分为2组(平均SSD = 92.9 cm)。在前两个疗程内进行体内测量。结果:SSD评估的重现性是在2个馏分之间的12个光束上进行的。在0.5 cm的公差范围内,定位误差为33%(4/12)。 2个TPS的同意率为75%(30/40)。结论:体内剂量测定法是评估患者位置变化和TPS剂量计算的非常有趣的工具。