首页> 外文期刊>Knee surgery, sports traumatology, arthroscopy: official journal of the ESSKA >A biomechanical comparison of the primary stability of two minimally invasive techniques for repair of ruptured Achilles tendon
【24h】

A biomechanical comparison of the primary stability of two minimally invasive techniques for repair of ruptured Achilles tendon

机译:两种微创技术修复跟腱断裂的主要稳定性的生物力学比较

获取原文
获取原文并翻译 | 示例

摘要

Purpose: To compare the primary stability of two minimally invasive procedures of Achilles tendon (AT) repair, namely a modified percutaneous repair of ruptured AT and the Achillon suture configuration. Methods: Eighteen (nine matched pairs) frozen ovine ATs were tenotomized 5 cm from the calcaneal insertion. In each pair, one tendon was randomly allocated to one of the two techniques: a modified percutaneous repair group and the Achillon device suture configuration. Specimens were tested performing an unidirectional tensile load to failure using a servo-hydraulic testing device (MTS Systems, Eden Prairie, MN, USA), controlled by an electropulse e3000 INSTRON machine (Instron Ltd, Buckinghamshire, UK). The tendons were then loaded to failure at a rate of 10 mm/s. Results: Two of the nine pairs of specimens were discarded because one specimen for each of the pair pulled out of the pneumatic clamp during mechanical testing. The remaining seven matched pairs were successfully tested. There were no differences in mean strength, mean maximum load, mean failure elongation, tension, stiffness and mode of failure between the two groups. Conclusion: The Achillon-like configuration and the modified percutaneous repair of ruptured AT provided similar biomechanical performance.
机译:目的:比较两种跟腱修复术的微创手术的主要稳定性,即改良的经皮AT修复破裂的跟腱缝合术和Achillon缝合结构。方法:从跟骨插入处5 cm处切入18个(九对匹配的)冷冻绵羊AT。在每对中,将一根腱随机分配给以下两种技术之一:改良的经皮修复组和Achillon装置缝合线配置。使用伺服液压测试设备(MTS Systems,美国明尼苏达州伊甸草原,MTS Systems),通过电脉冲e3000 INSTRON机器(英国白金汉郡,英斯特朗有限公司)控制,对样品进行了单向拉伸破坏测试。然后将腱以10 mm / s的速度加载至破坏。结果:丢弃了9对样品中的2对,因为在机械测试过程中,每对样品中有1个从气动夹具中拉出。其余的七对匹配已成功测试。两组之间的平均强度,平均最大负荷,平均破坏伸长率,张力,刚度和破坏模式没有差异。结论:类似于Achillon的配置和改良的经皮AT修复破裂的AT提供了相似的生物力学性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号