The output of electromechanical pressure gauges depends on not only the currently applied pressure, but also the pressurization history. Thus, the calibration results of gauges are affected by the pressurization procedure. In this paper, among several important factors influencing the results, we report the effects of the interval between the calibration cycles and the effects of the preliminary pressurizations. In order to quantitatively evaluate these effects, we developed a fully automated system that uses a pressure balance to calibrate pressure gauges. Subsequently, gauges containing quartz Bourdon-type pressure transducers were calibrated in a stepwise manner for pressures between 10 MPa and 100 MPa. The typical standard deviation of the data over three cycles was reduced to a few parts per million (ppm). The interval between the calibration cycles, which ranges from zero to more than 12 h, exerts a strong influence on the results in the process of increasing the pressure, where at 10 MPa the maximum difference between the results was approximately 40 ppm. The preliminary pressurization immediately before the calibration cycle reduces the effects of the interval on the results in certain cases. However, in turn, the influence of the waiting time between the preliminary pressurization and the main calibration cycle becomes strong. In the present paper, we outline several possible measures for obtaining calibration results with high reproducibility.
机译:机电压力表的输出不仅取决于当前施加的压力,还取决于增压历史。因此,压力计的校准结果会受到压力程序的影响。在本文中,在影响结果的几个重要因素中,我们报告了校准周期之间的间隔和预加压的影响。为了定量评估这些影响,我们开发了一种全自动系统,该系统使用压力平衡来校准压力表。随后,逐步将包含石英波登型压力传感器的压力表校准为10 MPa至100 MPa之间的压力。三个周期内数据的典型标准偏差降低到百万分之几(ppm)。校准周期之间的时间间隔(从零到大于12小时)对压力增加过程中的结果有很大影响,其中在10 MPa时,结果之间的最大差约为40 ppm。在某些情况下,即将在校准周期之前进行初步加压可减小间隔对结果的影响。但是,预加压和主校准周期之间的等待时间的影响会变强。在本文中,我们概述了获得具有高重现性的校准结果的几种可能的措施。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号