首页> 外文期刊>Measurement Science & Technology >The measurement of tiny dew droplets at the initial deposition stage and dew point using a phase-shift interference microscope
【24h】

The measurement of tiny dew droplets at the initial deposition stage and dew point using a phase-shift interference microscope

机译:使用相移干涉显微镜在初始沉积阶段和露点测量微小的露滴

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The shape of a dew droplet deposited on the mirror surface of a copper plate was measured accurately using an interference microscope that employed a phase-shift technique. The microscope was constructed by adding a piezoelectric transducer to an interference microscope. A simple method that uses a conventional speaker horn and an optical fibre cable was used to depress any speckle noise. The shape of a dew droplet deposited at dew point on the plate surface with average roughness of 0.1 μm was measured with an accuracy of ±3 nm. The mass of a tiny dew droplet could be determined from the volume of its shape and was of the order of 10~(-9) g. The total mass of a dew droplet deposited per unit area and the deposition velocity under a gentle wind were also obtained in a similar way. The total mass was of the order of 10~(-5) g cm~(-2) at the beginning of deposition and the deposition velocity ranged from 1×10~(-5) to 6×10~(-5) g cm~(-2) min~(-1) at room temperature.
机译:使用采用相移技术的干涉显微镜准确地测量沉积在铜板的镜面上的露珠的形状。通过在干涉显微镜上增加压电换能器来构造显微镜。使用常规扬声器喇叭和光纤电缆的一种简单方法可以抑制任何斑点噪声。以±3nm的精度测量以平均粗糙度为0.1μm在露点上沉积在板表面上的露滴的形状。从其形状的体积可以确定一个微小的露滴的质量,其大小约为10〜(-9)g。也以类似的方式获得了单位面积上沉积的露滴的总质量和在微风下的沉积速度。沉积开始时总质量为10〜(-5)g cm〜(-2),沉积速度为1×10〜(-5)至6×10〜(-5)g室温下cm〜(-2)min〜(-1)

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号